The influence of breathing system filters on paediatric capnography.

J G Hardman1, R P Mahajan, J Curran.   

Abstract

Breathing system filters are in common use during paediatric anaesthesia. Expired gas sampling from the patient side of these filters may contaminate and saturate the sampling line, while sampling from the machine side may cause underestimation of end-tidal carbon dioxide (PECO 2). The aim of this investigation was to elucidate the degree of underestimation of PECO 2 induced by sampling from the machine side of a breathing system filter. Ten spontaneously breathing children and ten children receiving mechanical ventilation under general anaesthesia were studied. PECO 2 was higher at the patient side of the filter in both ventilated and spontaneously breathing groups (P<0.002 for each). The bias in measuring at the machine side of the filter was significantly greater in the spontaneously breathing children as compared with the mechanically ventilated children (-1.8 vs -0.7 kPa; P<0.004).

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10712713

Source DB:  PubMed          Journal:  Paediatr Anaesth        ISSN: 1155-5645            Impact factor:   2.556


  3 in total

1.  Evaluation the effect of breathing filters on end-tidal carbon dioxide during inferior abdominal surgery in infants and changes of tidal volume and respiratory rate needs for preventing of increasing end-tidal carbon dioxide.

Authors:  Parvin Sajedi; Mohsen Abooei; Amir Shafa; Mahboobeh Karbalaei; Atefeh Babaei
Journal:  J Res Med Sci       Date:  2016-11-07       Impact factor: 1.852

2.  Comparison of mainstream end tidal carbon dioxide on Y-piece side versus patient side of heat and moisture exchanger filters in critically ill adult patients: a prospective observational study.

Authors:  Satoshi Tamashiro; Izumi Nakayama; Koichiro Gibo; Junichi Izawa
Journal:  J Clin Monit Comput       Date:  2022-08-03       Impact factor: 1.977

3.  Time course of end-tidal desflurane concentration during delivery and elimination according to the type and location of filters in a semi-closed circuit system.

Authors:  Sung-Wook Park; Jun-Young Chung; Eun-Jin Moon; Young-Jin Kang; Jong-Man Kang
Journal:  Korean J Anesthesiol       Date:  2015-07-28
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.