Literature DB >> 10711288

Stereotaxic core needle biopsy of breast microcalcifications obtained using a standard mammography table with an add-on unit.

S E Ward1, D H Taves, L I McCurdy.   

Abstract

OBJECTIVE: To demonstrate the reliability of stereotaxic biopsy of indeterminate microcalcifications using a standard mammography table with an add-on unit.
METHODS: In 121 cases of indeterminate microcalcifications, core biopsy was performed using a standard mammography table with an add-on stereotaxic unit. Microcalcifications were identified on radiography of core specimens.
RESULTS: Microcalcifications and a definitive histologic diagnosis were obtained in 112 core biopsies (92.6%), with no significant complications. In 23 lesions frank malignancy was diagnosed, and all of these diagnoses were confirmed on surgery. Pathologic examination suggested carcinoma in 4 lesions, and open biopsy confirmed malignancy in 3 of these cases. Four lesions showed atypical ductal hyperplasia. Benign disease was diagnosed in 81 lesions, of which 78 remained stable on mammographic follow-up (mean 16 months later) and 3 were subjected to surgical biopsy (of which 1 was malignant and 2 were benign). Nine cases were technically unsatisfactory because microcalcifications were not sampled.
CONCLUSION: Stereotaxic core biopsy performed with an add-on unit is a safe and reliable technique for biopsy of indeterminate microcalcifications. For successful biopsy, microcalcifications must be harvested. Pathologic results should be correlated with mammographic findings. The accuracy rate compares favourably with results reported using prone biopsy tables. In an era of cost containment, this alternative to prone biopsy tables could result in significant savings in terms of capital investment and use of hospital rooms. In this study, surgical biopsy could have been avoided in 64.5% of cases.

Entities:  

Mesh:

Year:  2000        PMID: 10711288

Source DB:  PubMed          Journal:  Can Assoc Radiol J        ISSN: 0846-5371            Impact factor:   2.248


  1 in total

1.  Randomized controlled trial of stereotactic 11-G vacuum-assisted core biopsy for the diagnosis and management of mammographic microcalcification.

Authors:  Sara M Bundred; Anthony J Maxwell; Julie Morris; Yit Y Lim; Md Janick Harake; Sigrid Whiteside; Nigel J Bundred
Journal:  Br J Radiol       Date:  2015-12-14       Impact factor: 3.039

  1 in total