Literature DB >> 10706828

Endovascular healing is inadequate for fixation of Dacron stent-grafts in human aortoiliac vessels.

M Malina1, J Brunkwall, K Ivancev, J Jönsson, J Malina, B Lindblad.   

Abstract

BACKGROUND: migration and kinking of stent-grafts can occur late after endovascular aneurysm repair. It is unknown if endovascular grafts incorporate enough to be permanently anchored. In this report, healing of aortic stent-grafts was assessed in humans. PATIENTS AND METHODS: we retrieved 23 Dacron stent-grafts from patients treated for an aortic aneurysm since 1993. Twelve stent-grafts were explanted at late conversion to open repair and 11 at autopsy. The deaths were unrelated to graft fixation. The median age of the patients was 74 years (IQR 55-84 years) and the grafts were explanted 9 months (1-31 months) after insertion. Microscopic slides were prepared by conventional techniques or by cutting and grinding arterial specimens embedded in plastic with the stent-grafts in situ.
RESULTS: the stent-grafts detached readily from the native arteries at surgery or autopsy, except when the stents had hooks or barbs which engaged the vessel wall. A space filled with poorly organised blood components persisted between the graft and the aortic wall 2.5 years after implantation. No firm incorporation of the grafts was observed proximally in the aneurysm neck or distally in the iliac segment. A friable neo-intimal layer covered parts of the luminal aspect of the grafts.
CONCLUSIONS: endovascular healing provides poor fixation of Dacron stent-grafts in humans. At present, fixation relies on the mechanical properties of the stent-grafts. Copyright 2000 Harcourt Publishers Ltd.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10706828     DOI: 10.1053/ejvs.1999.0867

Source DB:  PubMed          Journal:  Eur J Vasc Endovasc Surg        ISSN: 1078-5884            Impact factor:   7.069


  5 in total

1.  Patient demographics and cardiovascular risk factors differentially influence geometric remodeling of the aorta compared with the peripheral arteries.

Authors:  Dimitrios Miserlis; Peter Adamson; Micah Adamson; Thomas Knowles; Jamil Neme; Alexey Kamenskiy; Panagiotis Koutakis; Nicholas Phillips; Iraklis Pipinos; Jason MacTaggart
Journal:  Surgery       Date:  2015-06-19       Impact factor: 3.982

2.  Thoracic endovascular aortic repair with homemade stent grafts: early and mid-term outcomes.

Authors:  Seiji Onitsuka; Atsuhisa Tanaka; Hidetoshi Akashi; Keiichi Akaiwa; Keita Mikasa; Kurando Kanaya; Yusuke Shintani; Shinichi Hiromatsu; Shigeaki Aoyagi
Journal:  Surg Today       Date:  2011-07-20       Impact factor: 2.549

3.  Complementary roles of open and hybrid approaches to thoracoabdominal aortic aneurysm repair.

Authors:  Ehsan Benrashid; Hanghang Wang; Nicholas D Andersen; Jeffrey E Keenan; Richard L McCann; G Chad Hughes
Journal:  J Vasc Surg       Date:  2016-07-18       Impact factor: 4.268

Review 4.  A sustained-release drug-delivery system of synthetic prostacyclin agonist, ONO-1301SR: a new reagent to enhance cardiac tissue salvage and/or regeneration in the damaged heart.

Authors:  Satsuki Fukushima; Shigeru Miyagawa; Yoshiki Sakai; Yoshiki Sawa
Journal:  Heart Fail Rev       Date:  2015-07       Impact factor: 4.214

5.  Aortic dissection occurring 18 months after successful endovascular repair in an anatomically difficult case of abdominal aortic aneurysm.

Authors:  Satoshi Yamamoto; Katsuyuki Hoshina; Yutaka Takazawa; Hiroyuki Okamoto; Kunihiro Shigematsu; Tetsuro Miyata; Toshiaki Watanabe
Journal:  Case Rep Vasc Med       Date:  2013-11-04
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.