Literature DB >> 10701510

Dose rate calculations around 192Ir brachytherapy sources using a Sievert integration model.

P Karaiskos1, A Angelopoulos, P Baras, H Rozaki-Mavrouli, P Sandilos, L Vlachos, L Sakelliou.   

Abstract

The classical Sievert integral method is a valuable tool for dose rate calculations around brachytherapy sources, combining simplicity with reasonable computational times. However, its accuracy in predicting dose rate anisotropy around 192Ir brachytherapy sources has been repeatedly put into question. In this work, we used a primary and scatter separation technique to improve an existing modification of the Sievert integral (Williamson's isotropic scatter model) that determines dose rate anisotropy around commercially available 192Ir brachytherapy sources. The proposed Sievert formalism provides increased accuracy while maintaining the simplicity and computational time efficiency of the Sievert integral method. To describe transmission within the materials encountered, the formalism makes use of narrow beam attenuation coefficients which can be directly and easily calculated from the initially emitted 192Ir spectrum. The other numerical parameters required for its implementation, once calculated with the aid of our home-made Monte Carlo simulation code, can be used for any 192Ir source design. Calculations of dose rate and anisotropy functions with the proposed Sievert expression, around commonly used 192Ir high dose rate sources and other 192Ir elongated source designs, are in good agreement with corresponding accurate Monte Carlo results which have been reported by our group and other authors.

Mesh:

Substances:

Year:  2000        PMID: 10701510     DOI: 10.1088/0031-9155/45/2/309

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  4 in total

1.  Monte Carlo dose calculation of GZP6 (60)Co stepping source based on a matrix shift technique.

Authors:  Mohammad Taghi Bahreyni Toossi; Malihe Abdollahi; Mahdi Ghorbani
Journal:  Rep Pract Oncol Radiother       Date:  2010-12-21

2.  Using matrix summation method for three dimensional dose calculation in brachytherapy.

Authors:  Mahmoud Zibandeh-Gorji; Ali Asghar Mowlavi; Saeed Mohammadi
Journal:  Rep Pract Oncol Radiother       Date:  2012-02-09

3.  Current state of the art brachytherapy treatment planning dosimetry algorithms.

Authors:  P Papagiannis; E Pantelis; P Karaiskos
Journal:  Br J Radiol       Date:  2014-07-16       Impact factor: 3.039

4.  Intensity Modulated High Dose Rate (HDR) Brachytherapy Using Patient Specific 3D Metal Printed Applicators: Proof of Concept.

Authors:  James J Sohn; Mitchell Polizzi; Sang-Won Kang; Woo-Hyeong Ko; Yong-Hyun Cho; Keun-Yong Eom; Jin-Beom Chung
Journal:  Front Oncol       Date:  2022-02-10       Impact factor: 5.738

  4 in total