Literature DB >> 10700788

Improving the shape and compliance characteristics of a high-volume, low-pressure cuff improves tracheal seal.

P J Young1, M C Blunt.   

Abstract

A prototype design of a compliant latex, high-volume, low-pressure cuffed tracheal tube cuff (CHVLP) was compared with the Mallinckrodt Hi-Lo, Sheridan preformed and Portex Profile high-volume, low-pressure (HVLP) cuffed tracheal tubes for leakage of dye placed above the cuff in a benchtop mechanical ventilation model and in five isolated pig tracheas. There was no leakage in the ventilation model or in the pig tracheas with the prototype CHVLP. There was rapid leakage in the ventilation model and in all the pig tracheas for the Mallinckrodt Hi-Lo, the Sheridan preformed and the Portex Profile cuffs. This benchtop study suggests that improved HVLP cuff compliance characteristics may be beneficial in the prevention of leakage of fluid to the lungs known to occur with HVLP cuffs.

Entities:  

Mesh:

Year:  1999        PMID: 10700788     DOI: 10.1093/bja/83.6.887

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  5 in total

1.  Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube.

Authors:  Alexander Dullenkopf; Andreas Gerber; Markus Weiss
Journal:  Intensive Care Med       Date:  2003-08-16       Impact factor: 17.440

2.  Tracheal fluid leakage in benchtop trials: comparison of static versus dynamic ventilation model with and without lubrication.

Authors:  Mital H Dave; Nelly Koepfer; Caveh Madjdpour; Angela Frotzler; Markus Weiss
Journal:  J Anesth       Date:  2010-02-23       Impact factor: 2.078

3.  The effect of tracheal tube size on air leak around the cuffs.

Authors:  Jin-Young Hwang; Sang-Hyun Park; Sung-Hee Han; Seong-Joo Park; Soo-Kyung Park; Jin-Hee Kim
Journal:  Korean J Anesthesiol       Date:  2011-07-21

4.  The tracheal tube: gateway to ventilator-associated pneumonia.

Authors:  Parjam S Zolfaghari; Duncan L A Wyncoll
Journal:  Crit Care       Date:  2011-09-29       Impact factor: 9.097

5.  Effect of tracheal suctioning on aspiration past the tracheal tube cuff in mechanically ventilated patients.

Authors:  Pascal Beuret; Bénédicte Philippon; Xavier Fabre; Mahmoud Kaaki
Journal:  Ann Intensive Care       Date:  2012-11-07       Impact factor: 6.925

  5 in total