Literature DB >> 10699702

A case report of difficult ventilation with the Combitube - valve-like upper airway obstruction confirmed by fibreoptic visualisation.

G Jaehnichen1, N Golecki, M D Lipp.   

Abstract

This case report describes difficulty with ventilation because of valve-like upper airway obstruction by aryepiglottic folds after uncomplicated insertion of a Combitube in a 30-year-old female patient. After correct (oesophageal) placement increased ventilation pressures occurred and a fibreoptic device was used to investigate the cause. Valve-like obstruction was discovered and subsequently observed during controlled ventilation. After removal of the Combitube and mask ventilation no valve mechanism was seen. This effect appeared to be due to an increased air stream caused by the obstruction of seven out of eight Combitube perforations.

Entities:  

Mesh:

Year:  2000        PMID: 10699702     DOI: 10.1016/s0300-9572(99)00162-8

Source DB:  PubMed          Journal:  Resuscitation        ISSN: 0300-9572            Impact factor:   5.262


  1 in total

1.  Crisis management during anaesthesia: difficult intubation.

Authors:  A D Paix; J A Williamson; W B Runciman
Journal:  Qual Saf Health Care       Date:  2005-06