Literature DB >> 10685297

Sound transmission into incubators in the neonatal intensive care unit.

A Robertson1, C Cooper-Peel, P Vos.   

Abstract

OBJECTIVE: To measure the attenuation of sound by modern incubators. STUDY
DESIGN: LEQ, LMAX, LPEAK, and frequency distribution were measured simultaneously inside and outside two recent model incubators.
RESULTS: The attenuation of sound (outside minus inside) was 15 to 18 dBA with the motor off and 4 to 8 dBA with the motor on. There was a significant difference between incubators in their attenuation of sound. Octave band analysis showed attenuation in frequency bands of > 31.5 Hz with the motor off. With the motor on, the sound level inside the incubator was higher than outside at frequency bands of < 250 Hz.
CONCLUSION: Caring for infants inside modern incubators reduces "averaged" sound exposure to levels near those recommended for the neonatal intensive care unit. Lower frequency sounds are louder inside the incubator and arise from the incubator motor.

Entities:  

Mesh:

Year:  1999        PMID: 10685297     DOI: 10.1038/sj.jp.7200258

Source DB:  PubMed          Journal:  J Perinatol        ISSN: 0743-8346            Impact factor:   2.521


  2 in total

1.  Spectral analysis of noise in the neonatal intensive care unit.

Authors:  M D Livera; B Priya; A Ramesh; P N Suman Rao; V Srilakshmi; M Nagapoornima; A G Ramakrishnan; M Dominic
Journal:  Indian J Pediatr       Date:  2008-03       Impact factor: 1.967

2.  NICU Environment : Can we be Ignorant?

Authors:  Mng Nair; Girish Gupta; S K Jatana
Journal:  Med J Armed Forces India       Date:  2011-07-21
  2 in total