Literature DB >> 10683791

Incision width after phacoemulsification with foldable intraocular lens implantation.

N Mamalis1.   

Abstract

PURPOSE: To determine the incision size after insertion of foldable intraocular lenses (IOLs) using both a forceps and injectors.
SETTING: Intermountain Ocular Research Center, Moran Eye Center, University of Utah, Salt Lake City, Utah, USA.
METHODS: One hundred patients had phacoemulsification through a clear corneal wound. The incision width was measured before and after IOL implantation. A 3-piece silicone IOL and a 3-piece acrylic IOL (both with an optic diameter of 5.5 mm) were inserted using a forceps. A plate-haptic silicone lens and a 3-piece silicone lens with a 6.0 mm diameter optic were inserted using an injector.
RESULTS: The wound size in the group with the 3-piece silicone lens inserted with a forceps enlarged 4.4% (3.23 to 3.38 mm) in the low-power IOL group (15.0 to 20.0 diopters [D]) and 6.2% (3.24 to 3.44 mm) in the high-power IOL group (20.5 to 25.0 D). Similarly, the acrylic IOL wound enlarged 5% (3.21 to 3.37 mm) in the low-power IOL group and 6% (3.25 to 3.44 mm) in the high-power IOL group when a forceps was used. The 3-piece silicone and plate silicone lenses inserted using an injector enlarged the wound 3.2% and 3.3% (3.02 to 3.11 mm and 3.05 to 3.15 mm), respectively. There was no difference in the wound size with higher IOL powers in eyes with injected lenses.
CONCLUSIONS: Clear corneal incisions enlarge after insertion of foldable IOLs in a predictable manner, with a forceps-inserted IOL enlarging the wound diameter more than lenses inserted with an injector. The forceps-inserted lens group also showed a difference in wound size related to IOL power.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10683791     DOI: 10.1016/s0886-3350(99)00346-6

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  2 in total

Review 1.  [Incisions for biaxial and coaxial microincision cataract surgery].

Authors:  M Müller; T Kohnen
Journal:  Ophthalmologe       Date:  2010-02       Impact factor: 1.059

2.  Comparison of surgically induced astigmatisms after clear corneal incisions of different sizes.

Authors:  Sung Chur Moon; Tarek Mohamed; I Howard Fine
Journal:  Korean J Ophthalmol       Date:  2007-03
  2 in total