Literature DB >> 10678384

First clinical experience with a mechanical valve with silver coating.

A Brutel de la Riviere1, K M Dossche, D E Birnbaum, R Hacker.   

Abstract

BACKGROUND AND AIM OF THE STUDY: The interface between the annulus and sewing cuff is the infectious center of prosthetic valve endocarditis (PVE). To decrease the incidence of PVE, the sewing cuff of the St. Jude Medical (SJM) mechanical heart valve was permanently coated with elemental silver (Silzone coating). In vitro data have supported the antimicrobial efficacy of this coating.
METHODS: To study any adverse effects of the silver coating in humans, serum silver levels were determined (by graphite furnace atomic absorption spectrometry) before (baseline) and at five intervals after operation: day 1, day 3, discharge, one month, and two months. Between January and August 1997, 38 patients (71% males) underwent surgical implant of a SJM Masters Series valve with Silzone coating for the aortic valve (n = 29), mitral valve (n = 6), or both valves (n = 3). Five patients (13%) underwent concomitant procedures. Two patients (5%) presented with native active valve endocarditis.
RESULTS: There was no hospital mortality or valve-related hospital morbidity. Blood silver concentrations peaked shortly after surgery and then decreased during the postoperative period. Average levels were consistently below 4 parts per billion (ppb). Levels below 10 ppb are considered normal. Follow up was 95% complete. There were no recurrent or new cases of endocarditis.
CONCLUSION: These clinical data indicate that the SJM Masters Series valve with Silzone coating performs well. No adverse effects of the silver coating could be detected, and there were no valve-related complications.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10678384

Source DB:  PubMed          Journal:  J Heart Valve Dis        ISSN: 0966-8519


  17 in total

Review 1.  Silver-coated megaprostheses: review of the literature.

Authors:  Tom Schmidt-Braekling; Arne Streitbuerger; Georg Gosheger; Friedrich Boettner; Markus Nottrott; Helmut Ahrens; Ralf Dieckmann; Wiebke Guder; Dimosthenis Andreou; Gregor Hauschild; Burkhard Moellenbeck; Wenzel Waldstein; Jendrik Hardes
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-03-06

2.  Silver nanoparticle deposited implants to treat osteomyelitis.

Authors:  Samit Kumar Nandi; Anish Shivaram; Susmita Bose; Amit Bandyopadhyay
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2017-05-15       Impact factor: 3.368

3.  Mechanical degradation of TiO2 nanotubes with and without nanoparticulate silver coating.

Authors:  Anish Shivaram; Susmita Bose; Amit Bandyopadhyay
Journal:  J Mech Behav Biomed Mater       Date:  2016-03-03

Review 4.  Silver-coated megaprosthesis in prevention and treatment of peri-prosthetic infections: a systematic review and meta-analysis about efficacy and toxicity in primary and revision surgery.

Authors:  Michele Fiore; Andrea Sambri; Riccardo Zucchini; Claudio Giannini; Davide Maria Donati; Massimiliano De Paolis
Journal:  Eur J Orthop Surg Traumatol       Date:  2020-09-05

5.  Understanding long-term silver release from surface modified porous titanium implants.

Authors:  Anish Shivaram; Susmita Bose; Amit Bandyopadhyay
Journal:  Acta Biomater       Date:  2017-05-29       Impact factor: 8.947

Review 6.  Use of silver in the prevention and treatment of infections: silver review.

Authors:  Amani D Politano; Kristin T Campbell; Laura H Rosenberger; Robert G Sawyer
Journal:  Surg Infect (Larchmt)       Date:  2013-02-28       Impact factor: 2.150

7.  Interactions of neutrophils with silver-coated vascular polyester grafts.

Authors:  Joerg Tautenhahn; Frank Meyer; Thomas Buerger; Uwe Schmidt; Hans Lippert; Wolfgang Koenig; Brigitte Koenig
Journal:  Langenbecks Arch Surg       Date:  2008-12-02       Impact factor: 3.445

8.  In vitro antimicrobial properties of silver-polysaccharide coatings on porous fiber-reinforced composites for bone implants.

Authors:  Sara Nganga; Andrea Travan; Eleonora Marsich; Ivan Donati; Eva Söderling; Niko Moritz; Sergio Paoletti; Pekka K Vallittu
Journal:  J Mater Sci Mater Med       Date:  2013-08-07       Impact factor: 3.896

9.  Measurement of the silver ion concentration in wound fluids after implantation of silver-coated megaprostheses: correlation with the clinical outcome.

Authors:  B Hussmann; I Johann; M D Kauther; S Landgraeber; Marcus Jäger; S Lendemans
Journal:  Biomed Res Int       Date:  2013-05-29       Impact factor: 3.411

10.  The influence of elementary silver versus titanium on osteoblasts behaviour in vitro using human osteosarcoma cell lines.

Authors:  Jendrik Hardes; Arne Streitburger; Helmut Ahrens; Thomas Nusselt; Carsten Gebert; Winfried Winkelmann; Achim Battmann; Georg Gosheger
Journal:  Sarcoma       Date:  2007
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.