V F Nascimento Filho1, V H Poblete, P S Parreira, E Matsumoto, S M Simabuco, E P Espinoza, A A Navarro.
Abstract
An X-ray tube with a Mo target and Zr filter, operated at 45 kV/20 mA, was used to excite samples (5 microL deposited on a quartz support) and the total reflection angle condition was obtained with a double reflector module built with two 10-cm-long 7-mm-thick quartz crystals placed 50 microns apart. A high-resolution spectrometer based on a Si(Li) detector coupled to a multichannel analyzer was used for X-ray detection and the spectra were interpreted with the AXIL software. The system was calibrated with standard chemical solutions containing Cr, Fe, Cu, Zn, and Pb, and Y was used as an internal standard to correct eventual geometric errors and high-voltage instabilities of the X-ray generator. The limits of detection were 19, 9, 5, and 4 ng/mL for Cr, Fe, Cu, and Zn, respectively, analyzed through characteristic K alpha X-rays, and 7 ng/mL for Pb, through L alpha X-rays, considering 50 microL samples deposited and dried on a quartz support, to be excited/detected for 1000 s.Entities:
Mesh:
Substances:
Year: 1999 PMID: 10676518 DOI: 10.1007/bf02784230