| Literature DB >> 1067140 |
A C Enright, R L Moore, F L Parney.
Abstract
Fifty Bain circuits were tested for contamination following single patient use. A contamination rate of 8 per cent was found. Two methods of resterilization were examined. Sterilization by activated glutaraldehyde proved unsatisfactory, while ethylene oxide sterilization was found to be an effective method which produced no detectable adverse physical or chemical alterations of the circuits.Entities:
Mesh:
Substances:
Year: 1976 PMID: 1067140 DOI: 10.1007/bf03005983
Source DB: PubMed Journal: Can Anaesth Soc J ISSN: 0008-2856