Literature DB >> 10667664

Delineation of simulated vascular stenosis with Gd-DTPA-enhanced 3D gradient echo MR angiography: an experimental study.

K Mitsuzaki1, Y Yamashita, M Onomichi, T Tsuchigame, M Takahashi.   

Abstract

PURPOSE: The purpose of this experimental study was to evaluate the influence of contrast material concentration and flow velocity on pulsatile flow in Gd-DTPA-enhanced 3D gradient echo MR angiographic sequence.
METHOD: In vivo flow experiments were performed in Plexiglas phantoms with artificial stenosis (50% stenotic ratio and 20 mm stenotic length) attached to a cardiac pump that generated physiological pulsatile flow similar to that of the bloodstream in a closed circuit. We used a steady-state gradient echo sequence with different TEs (6, 3, and 1.4 ms). A TR of 15 ms was used for all parameters. The concentration of Gd-DTPA varied from 0 to 2.0 mmol/L and flow velocities from 25 to 80 cm/s. We measured the degree of stenosis and length of stenosis in comparison with the actual values.
RESULTS: The degree and length of stenosis on 3D gradient echo MR angiographic images were markedly influenced by the velocity of the flow and concentration of Gd-DTPA. The degree of stenosis was overestimated when the flow was fast or when the concentration of Gd-DTPA was low. When the concentration of Gd-DTPA was low, stenosis was elongated. These effects were less prominent on short TE (1.4 ms) sequence.
CONCLUSION: The stenotic lesions were markedly overestimated on MR angiographic images obtained with Gd-DTPA-enhanced fast 3D gradient echo sequence. Spin dephasing can be compensated for almost entirely by a high concentration of Gd-DTPA and/or a short TE sequence.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10667664     DOI: 10.1097/00004728-200001000-00016

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  1 in total

1.  Evaluation of the intracranial dural sinuses with a 3D contrast-enhanced MP-RAGE sequence: prospective comparison with 2D-TOF MR venography and digital subtraction angiography.

Authors:  L Liang; Y Korogi; T Sugahara; M Onomichi; Y Shigematsu; D Yang; M Kitajima; Y Hiai; M Takahashi
Journal:  AJNR Am J Neuroradiol       Date:  2001-03       Impact factor: 3.825