Virtual computed tomography colonoscopy: artifacts, image quality and radiation dose load in a cadaver study.

P Springer1, B Stöhr, S M Giacomuzzi, G Bodner, A Klingler, W Jaschke, D zur Nedden.   

Abstract

The purpose of our study was to evaluate the interdependency of spatial resolution, image reconstruction artifacts, and radiation doses in virtual CT colonoscopy by comparing various CT scanning protocols. A pig's colon with several artificial polypoid lesions was imaged after air insufflation with helical CT scanning using 1-, 3-, and 5-mm collimation, and pitch values varying from 1.0 to 3.0. Virtual endoscopic images and "fly through" sequences were calculated on a Sun Sparc 20 workstation (Navigator Software, GE Medical Systems, Milwaukee, Wis.). Several reconstruction artifacts as well as overall image quality were evaluated by three independent reviewers. In addition, radiation doses for the different CT protocols were measured as multiple-scan average dose using a 10-cm ion chamber and a standard Plexiglass body phantom. Generally, image quality and reconstruction artifacts were less affected by pitch values than by beam collimation. Thus, narrow beam collimation at higher pitch values (e. g. 3 mm/2.0) seems to be a reasonable compromise between quality of virtual endoscopic images and radiation dose load.

Entities:  

Mesh:

Year:  2000        PMID: 10663741     DOI: 10.1007/s003300050030

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  3 in total

1.  Spiral CT in gastric carcinoma: comparison with barium study, fiberoptic gastroscopy and histopathology.

Authors:  Feng Chen; Yi-Cheng Ni; Kai-Er Zheng; Sheng-Hong Ju; Jun Sun; Xi-Long Ou; Man-Hua Xu; Hao Zhang; Guy Marchal
Journal:  World J Gastroenterol       Date:  2003-07       Impact factor: 5.742

2.  Diagnostic contribution of virtual endoscopy in diseases of the upper airways.

Authors:  Süleyman Men; M Cenk Ecevit; I Topçu; Neslihan Kabakçi; Taner K Erdağ; Semih Sütay
Journal:  J Digit Imaging       Date:  2007-03       Impact factor: 4.056

3.  Computed Tomography Colonography Phantom: Construction, Validation and Literature Review.

Authors:  Lukas Lambert; Alena Lambertova; Jan Danes; Gabriela Grusova
Journal:  Iran J Radiol       Date:  2016-06-22       Impact factor: 0.212

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.