Abstract

A photon virtual source model was developed for simulating arbitrary, external beam, intensity distributions using the Monte Carlo method. The source model consists of a photon fluence grid composed of a matrix of square elements, located 25-cm downstream from the linear accelerator target. Each particle originating from the fluence map is characterized by the seven phase space parameters, position (x, y, z), direction (u, v, w), and energy. The map was reconstructed from fluence and energy spectra acquired by modeling components of the linear accelerator treatment head using the Monte Carlo code MCNP4B. The effect of contaminant electrons is accounted for by the use of a sub-source derived from a phase-space simulation of a 25-MV linac treatment head using the code BEAM. The BEAM sub-source was incorporated into the MCNP4B phase-space model and is sampled using a field-size dependent sampling ratio. A Gaussian blurring kernel is convolved with the photon fluence map to account for the finite focal spot size and scattering effects from structures such as the flattening filter and MLC leaves. Depth dose and profile source calculations for 6-MV and 25-MV photon beams, for 5 x 5 cm2, 10 x 10 cm2, and 15 x 15 cm2 field sizes, are in good agreement with measurement and are well within acceptability criteria suggested by the AAPM Task Group Report No. 53. Irregular field calculations compared with film measurement and with a 3-D pencil beam algorithm show that the source model is capable of accurately simulating arbitrary MLC fields.

Entities:  

Mesh:

Year:  2000        PMID: 10659754     DOI: 10.1118/1.598881

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  Monte Carlo simulation for Neptun 10 PC medical linear accelerator and calculations of output factor for electron beam.

Authors:  Mohammad Taghi Bahreyni Toossi; Mehdi Momennezhad; Seyed Mohammad Hashemi
Journal:  Rep Pract Oncol Radiother       Date:  2012-03-06

2.  Determination of radiotherapy X-ray spectra using a screen-film system.

Authors:  H M Garnica-Garza
Journal:  Med Biol Eng Comput       Date:  2008-09-09       Impact factor: 2.602

3.  A measurement-based generalized source model for Monte Carlo dose simulations of CT scans.

Authors:  Xin Ming; Yuanming Feng; Ransheng Liu; Chengwen Yang; Li Zhou; Hezheng Zhai; Jun Deng
Journal:  Phys Med Biol       Date:  2017-01-12       Impact factor: 3.609

Review 4.  Monte Carlo methods for device simulations in radiation therapy.

Authors:  Hyojun Park; Harald Paganetti; Jan Schuemann; Xun Jia; Chul Hee Min
Journal:  Phys Med Biol       Date:  2021-09-14       Impact factor: 4.174

5.  A virtual source model for Monte Carlo simulation of helical tomotherapy.

Authors:  Jiankui Yuan; Yi Rong; Quan Chen
Journal:  J Appl Clin Med Phys       Date:  2015-01-08       Impact factor: 2.102

6.  A single-source photon source model of a linear accelerator for Monte Carlo dose calculation.

Authors:  Obioma Nwankwo; Gerhard Glatting; Frederik Wenz; Jens Fleckenstein
Journal:  PLoS One       Date:  2017-09-08       Impact factor: 3.240

7.  Development of a virtual source model for Monte Carlo-based independent dose calculation for varian linac.

Authors:  James R Castle; Jingwei Duan; Xue Feng; Quan Chen
Journal:  J Appl Clin Med Phys       Date:  2022-02-09       Impact factor: 2.243

8.  Monte Carlo modeling of HD120 multileaf collimator on Varian TrueBeam linear accelerator for verification of 6X and 6X FFF VMAT SABR treatment plans.

Authors:  Alanah M Bergman; Ermias Gete; Cheryl Duzenli; Tony Teke
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102