Literature DB >> 10658697

A new device to limit extravasation during contrast-enhanced CT.

C C Powell1, J M Li, L Rodino, F A Anderson.   

Abstract

OBJECTIVE: The extravasation detection accessory (EDA) is designed for use during contrast-enhanced CT studies performed with a power injector. The EDA detects the changes in soft-tissue impedance that occur with enhanced extravasation and halts the further infusion of contrast material via a feedback circuit to the injector. We tested the sensitivity of this device in a model of contrast extravasation.
MATERIALS AND METHODS: Study subjects had an extravasation of 5% dextrose in water (nonionic contrast equivalent) in one arm and 0.9% sodium chloride solution (ionic contrast equivalent) in the other. An EDA was placed over the site of infusion and connected to a power injector. Injections were performed at 0.25 ml/sec (n = 40), 2.5 ml/sec (n = 62), or 5 ml/sec (n = 20).
RESULTS: At infusion rates of 2.5 and 5 ml/sec, the device halted the injector in every subject after an average volume of 12.5 +/- 1.6 ml was delivered. At 0.25 ml/sec, the device failed to halt the injector in 11 of 20 events. After reprogramming the algorithm, 10 more subjects were tested at the lowest injection rate. The device halted 18 of 20 extravasation events with an average volume of 3.7 +/- 0.5 ml.
CONCLUSION: In our model of contrast extravasation, the EDA halted a power injector with reliability and reproducibility before a large volume of contrast material was delivered. The sensitivity of the device approached, but did not reach, 100%. This device may serve to diminish the morbidity of extravasation events.

Entities:  

Mesh:

Year:  2000        PMID: 10658697     DOI: 10.2214/ajr.174.2.1740315

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  4 in total

1.  Method for performing cerebral perfusion-weighted MRI in neonates.

Authors:  Tarek Laswad; Pia Wintermark; Leonor Alamo; Adrien Moessinger; Reto Meuli; François Gudinchet
Journal:  Pediatr Radiol       Date:  2008-12-23

2.  Intravenous contrast medium extravasation: systematic review and updated ESUR Contrast Media Safety Committee Guidelines.

Authors:  Giles Roditi; Nadir Khan; Aart J van der Molen; Marie-France Bellin; Michele Bertolotto; Torkel Brismar; Jean-Michel Correas; Ilona A Dekkers; Remy W F Geenen; Gertraud Heinz-Peer; Andreas H Mahnken; Carlo C Quattrocchi; Alexander Radbruch; Peter Reimer; Laura Romanini; Fulvio Stacul; Henrik S Thomsen; Olivier Clément
Journal:  Eur Radiol       Date:  2022-02-17       Impact factor: 7.034

3.  Biceps Brachii Compartment Contrast Media Extravasation with Surgical Correlation.

Authors:  Felix S Chew; Carol A Boles; Christopher T Mattern
Journal:  Radiol Case Rep       Date:  2015-11-06

4.  Novel Conformal Skin Patch with Embedded Thin-Film Electrodes for Early Detection of Extravasation.

Authors:  Ruiqi Lim; Ramona B Damalerio; Choon Looi Bong; Swee Kim Tan; Ming-Yuan Cheng
Journal:  Sensors (Basel)       Date:  2021-05-14       Impact factor: 3.576

  4 in total