A prospective audit of pacemaker function, implant lifetime, and cause of death in the patient.

S K Suvarna1, R D Start, D I Tayler.   

Abstract

AIM: To audit prospectively the reasons for pacemaker implantation, the duration of the pacemaker use, the cause of death, and pacemaker function after removal from the patient.
METHODS: Pacemakers were removed at necropsy, or from the bodies of patients awaiting cremation, in three hospitals over a three year period. The cause of death was taken from the results of the necropsy or from the certified cause of death. Demographic data, including the time of implant and reasons for implantation, were checked. The pacemakers were analysed in terms of battery status, program, and output under a standard 470 ohm load.
RESULTS: 69 patients were studied. Average age at death was 78 and 80 years for men and women, respectively. The average duration since pacemaker implantation was 46 months. Eleven patients had necropsies, showing that three died from ischaemic heart disease, six from cardiomyopathy, one from an aortic aneurysm, and one from disseminated neoplasia. From the necropsy results and death certificates, the distribution of causes of death in the group as a whole were ischaemic heart disease (21), cardiomyopathy (8), cerebrovascular disease (11), neoplasia (11), chest infection/chronic obstructive airways disease (8), and other causes (10). In all cases the pacemaker box function was within normal limits.
CONCLUSIONS: Neither primary nor secondary pacemaker dysfunction was found. The study highlights the impact of arrhythmias in cardiomyopathy, and raises questions about the true role of ischaemic heart disease in these pacemaker requiring patients. The relatively short gap between pacemaker implantation and death requires further study.

Entities:  

Mesh:

Year:  1999        PMID: 10655989      PMCID: PMC501543          DOI: 10.1136/jcp.52.9.677

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  12 in total

1.  Experience with recycled cardiac pacemakers.

Authors:  K K Sethi; M Bhargava; N Pandit; J C Mohan; R Arora; S K Khanna; M Khalilullah
Journal:  Indian Heart J       Date:  1992 Mar-Apr

2.  A transistorized, self-contained, implantable pacemaker for the long-term correction of complete heart block.

Authors:  W M CHARDACK; A A GAGE; W GREATBATCH
Journal:  Surgery       Date:  1960-10       Impact factor: 3.982

3.  An intracardiac pacemaker for Stokes-Adams seizures.

Authors:  S FURMAN; J B SCHWEDEL
Journal:  N Engl J Med       Date:  1959-11-05       Impact factor: 91.245

4.  Survival and functional independence after implantation of a permanent pacemaker in octogenarians and nonagenarians. A population-based study.

Authors:  W K Shen; D L Hayes; S C Hammill; K R Bailey; D J Ballard; B J Gersh
Journal:  Ann Intern Med       Date:  1996-09-15       Impact factor: 25.391

5.  Autopsy findings with permanent pervenous pacemakers.

Authors:  S J Robboy; J W Harthorne; R C Leinbach; C A Sanders; W G Austen
Journal:  Circulation       Date:  1969-04       Impact factor: 29.690

Review 6.  Pacemaker malfunction.

Authors:  D L Hayes; R E Vlietstra
Journal:  Ann Intern Med       Date:  1993-10-15       Impact factor: 25.391

7.  Reuse of pacemaker.

Authors:  M Panja; C N Sarkar; S Kumar; A K Kar; S Mitra; D P Sinha; A Chatterjee; S Roy; N C Sarkar; B Majumder
Journal:  Indian Heart J       Date:  1996 Nov-Dec

8.  Survey of cardiac pacing in Canada (1993).

Authors:  B S Goldman; S Nishimura; C Lau
Journal:  Can J Cardiol       Date:  1996-06       Impact factor: 5.223

9.  Causes of death in patients with unipolar single chamber ventricular pacing: prevalence and circumstances in dependence on arrhythmias leading to pacemaker implantation.

Authors:  A V Mattioli; R Rossi; E Annicchiarico; G Mattioli
Journal:  Pacing Clin Electrophysiol       Date:  1995-01       Impact factor: 1.976

10.  Impact of internal audit on pacemaker prescription and the immediate costs of pacing in the northern region: towards implementation of the recommendations of the British Pacing and Electrophysiology Group.

Authors:  J P Mounsey; S G Ray; M J Griffith; R G Gold; R S Bexton
Journal:  Br Heart J       Date:  1994-04
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.