M B Esler1, D W Griffith, S R Wilson, L P Steele.
Abstract
We report the development of a method of carbon stable isotope ratio analysis based on 1-cm-1 resolution Fourier transform infrared (FT-IR) spectroscopy, deployable in both laboratory and field applications. We demonstrate the determination of the 13C/12C ratio of CO2 (i.e., delta 13CO2) in air with an analytical precision of the order of +/- 0.1/1000 (i.e., +/- 0.01%). The FT-IR method relies on calibration using synthetically calculated absorbance spectra and a multivariate calibration algorithm. The method requires no sample preparation other than optional drying of the sample and may be applied directly to ambient air samples containing approximately 350 mumol mol-1 CO2 (molar mixing ratio). It may also be applied to samples more concentrated in CO2, such as human breath, approximately 5% CO2. We demonstrate the utility of the technique to the analysis of delta 13CO2 in air during an experimental field campaign and to the laboratory-based analysis of human breath. A similar method could also be used to determine the H/D ratio in atmospheric water vapor.Entities:
Mesh:
Substances:
Year: 2000 PMID: 10655656 DOI: 10.1021/ac990563x
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986