Literature DB >> 10644108

MR hysterosalpingography: protocol development and refinement for simulating normal and abnormal fallopian tube patency--feasibility study with a phantom.

R E Frye1, S M Ascher, D Thomasson.   

Abstract

PURPOSE: To develop and refine a pulse sequence and protocol for testing the feasibility of magnetic resonance (MR) hysterosalpingography in a phantom model.
MATERIALS AND METHODS: A phantom simulating the uterus, fallopian tubes, and surrounding pelvic cavity was constructed. T2-weighted acquisition strategies-breath-hold fast spin-echo, rapid acquisition with relaxation enhancement (RARE), and haff-Fourier RARE-were refined to acquire sequential 70-mm coronal imaging volumes. Contrast agent was injected into the introducing catheter entering the os of the simulated uterus. Interacquisition interval, type of contrast agent (eg, sterile saline solution or water), and quantity of contrast agent (eg, 1-5 mL per acquisition) were varied. Digital image subtraction was used to enhance image quality. Images were qualitatively analyzed and rated good, fair, or poor for temporal resolution, spatial resolution, fallopian tube conspicuity, and free spill conspicuity. Once the technique was refine, the phantom was reconfigured to simulate unilateral and bilateral hydrosalpinx.
RESULTS: The RARE sequence with an 8-second interacquisition interval and a 5-mL interacquisition of sterile water produced good images of the simulated fallopian tubes and free spill. Depiction of unilateral and bilateral hydrosalpinx was also reliably demonstrated.
CONCLUSION: This study with a phantom model demonstrates the feasibility of MR hysterosalpingography to depict normal and diseased fallopian tubes.

Entities:  

Mesh:

Year:  2000        PMID: 10644108     DOI: 10.1148/radiology.214.1.r00ja42107

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  2 in total

1.  Conventional and magnetic resonance hysterosalpingography in assessing tubal patency-A comparative study.

Authors:  Devimeenal Jagannathan; Fouzal Hithaya
Journal:  Indian J Radiol Imaging       Date:  2019 Apr-Jun

2.  Comparison between X-ray-hysterosalpingography and 3 Tesla magnetic resonance-hysterosalpingography in the assessment of the tubal patency in the cause of female infertility.

Authors:  Cristina Pace; Renato Argirò; Luisa Casadei; Matteo Cesareni; Antonio Orlacchio
Journal:  Radiol Med       Date:  2022-09-27       Impact factor: 6.313

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.