Reliability, validity, and utility of various occlusal measurement methods and techniques.

K Baba1, Y Tsukiyama, G T Clark.   

Abstract

STATEMENT OF PROBLEM: The controversy continues regarding the efficacy of several commercially available devices that are used as aids in the diagnosis of occlusal abnormalities.
PURPOSE: This article reviews the reliability, validity, and utility of instruments that claim to detect occlusal interferences and abnormal vertical dimension of occlusion.
MATERIAL AND METHODS: Data, opinions, and technical information from 37 published articles were reviewed. Evidence for method reliability, validity, and utility was assessed.
RESULTS: Although occlusal contact detection devices can document the occlusal contact patterns on teeth, the basic in vivo testing of their reproducibility and validity has not been performed. Moreover, while EMG and jaw tracking systems can indeed measure jaw muscle activation and jaw position, no cost-benefit analysis of these devices has yet been conducted. One manufacturer suggests that these instruments be used in conjunction with an electrical muscle stimulation device to find a new resting jaw position that is more open vertically. This new, more open position has been inappropriately labeled as evidence of vertical dimension of occlusion over-closure.
CONCLUSION: None of the instruments reviewed can be said to be more than ancillary documentation devices and they have been inadequately tested for reliability and validity. Moreover, because scientifically acceptable disease definitions are not available for these 2 conditions, the issue of over-diagnosing becomes a substantial concern.

Entities:  

Mesh:

Year:  2000        PMID: 10633026     DOI: 10.1016/s0022-3913(00)70092-8

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  8 in total

1.  A survey on German dentists regarding the management of craniomandibular disorders.

Authors:  Michelle Alicia Ommerborn; Carolin Kollmann; Jörg Handschel; Rita Antonia Depprich; Hermann Lang; Wolfgang Hans-Michael Raab
Journal:  Clin Oral Investig       Date:  2009-05-14       Impact factor: 3.573

Review 2.  History of materials used for recording static and dynamic occlusal contact marks: a literature review.

Authors:  Ashu Sharma; G R Rahul; Soorya T Poduval; Karunakar Shetty; Bhawna Gupta; Varun Rajora
Journal:  J Clin Exp Dent       Date:  2013-02-01

3.  Differences between centric relation and maximum intercuspation as possible cause for development of temporomandibular disorder analyzed with T-scan III.

Authors:  Zana D Lila-Krasniqi; Kujtim Sh Shala; Teuta Pustina-Krasniqi; Teuta Bicaj; Linda J Dula; Ljuben Guguvčevski
Journal:  Eur J Dent       Date:  2015 Oct-Dec

4.  Differences between Subjective Balanced Occlusion and Measurements Reported With T-Scan III.

Authors:  Zana Lila-Krasniqi; Kujtim Shala; Teuta Pustina Krasniqi; Teuta Bicaj; Enis Ahmedi; Linda Dula; Arlinda Tmava Dragusha; Ljuben Guguvcevski
Journal:  Open Access Maced J Med Sci       Date:  2017-08-08

5.  Occlusal status of implant superstructures at mandibular first molar immediately after setting.

Authors:  Yukihiko Okada; Yuji Sato; Noboru Kitagawa; Keiichiro Uchida; Tokiko Osawa; Yoshiki Imamura; Mayumi Terazawa
Journal:  Int J Implant Dent       Date:  2015-07-04

6.  Effect of scan delay on measurements of an occlusal pressure sensitive film: An in-vitro study.

Authors:  Tong-Mei Wang; Yung-Hsiang Chang; Tsung-Chieh Yang; Li-Deh Lin
Journal:  J Dent Sci       Date:  2021-08-28       Impact factor: 2.080

7.  Occlusal Indicators Used in Dental Practice: A Survey Study.

Authors:  Tanya Bozhkova; Nina Musurlieva; Diyan Slavchev; Mariana Dimitrova; Sevda Rimalovska
Journal:  Biomed Res Int       Date:  2021-11-03       Impact factor: 3.411

8.  Reestablishment of Occlusal Vertical Dimension in Complete Denture Wearing in Two Stages.

Authors:  Danny Omar Mendoza Marin; Andressa Rosa Perin Leite; Norberto Martins de Oliveira Junior; Marco Antonio Compagnoni; Ana Carolina Pero; João Neudenir Arioli Filho
Journal:  Case Rep Dent       Date:  2015-10-26
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.