The "Bernese" frameless optical computer aided surgery system.

M Caversaccio1, R Bächler, K Lädrach, G Schroth, L P Nolte, R Häusler.   

Abstract

OBJECTIVE: We report on two years of clinical experience with a frameless Computer Aided Surgery system developed in Bern, Switzerland.
MATERIAL AND METHODS: Our navigation system is based on a preoperative computer-tomography (CT) scan (without markers) and an intraoperative optical tracking of head movements and of the surgical instruments. Using landmark and surface-based registration, the skull can be accurately correlated to the CT images. The three-dimensional positions of the surgical instruments, as well as the endoscopic images, are displayed in real time on a monitor.
RESULTS: In the last two years, 109 computer-aided interventions have been successfully performed: 89 on the anterior skull base/paranasal sinuses, 15 on the lateral skull base, and 5 minimally invasive procedures on other locations in the skull. No complications occurred. The practical accuracy on the cadaver skull is between 0.5 mm and 1.2 mm, and the clinical accuracy is between 0. 5 mm and 2 mm.
CONCLUSIONS: Our navigation system has proven its accuracy and usability. Surgeons feel very comfortable with the increased safety provided by the unequivocal identification of important anatomical structures. Copyright 1999 Wiley-Liss, Inc.

Mesh:

Year:  1999        PMID: 10631376     DOI: 10.1002/(SICI)1097-0150(1999)4:6<328::AID-IGS5>3.0.CO;2-Y

Source DB:  PubMed          Journal:  Comput Aided Surg        ISSN: 1092-9088


  2 in total

1.  On "Evaluation of the Contribution of CAS in Combination with the Subcranial/Subfrontal Approach in Anterior Skull Base Surgery" (Skull Base 2001;11:59-76).

Authors: 
Journal:  Skull Base       Date:  2002-02

2.  Feasibility of A-mode ultrasound based intraoperative registration in computer-aided orthopedic surgery: A simulation and experimental study.

Authors:  Kenan Niu; Jasper Homminga; Victor I Sluiter; André Sprengers; Nico Verdonschot
Journal:  PLoS One       Date:  2018-06-13       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.