Literature DB >> 10631375

Frameless stereotaxy and anterior cervical surgery.

C Bolger1, C Wigfield, T Melkent, K Smith.   

Abstract

OBJECTIVE: The role of a frameless image guidance system for anterior cervical surgery was evaluated in 40 patients with cervical spondylosis. An assessment was made of the challenges posed in achieving accurate three-dimensional image registration in the absence of multiple fixed anatomical landmarks, and a solution to this problem is outlined using a novel registration tracking device.
MATERIALS AND METHODS: Patients underwent preoperative CT scanning of the surgical field of interest, and the image data was reconstructed on a StealthStation (Sofamor Danek, Memphis, Tennessee, USA). An image guidance tracking device was attached to a modified Caspar retractor. Point matching and surface mapping were used for registration.
RESULTS: It was not possible to achieve accurate registration in the first 7 patients. In the next 5 patients, the time taken to achieve registration was prolonged. For 28 patients, registration was achieved in an acceptable time with an accuracy of 0.74 +/- 0.4 mm. Anatomical resection margins were accurately identified and 36 anterior cervical screws were precisely placed.
CONCLUSIONS: Frameless stereotaxy can now be applied to procedures involving the anterior cervical spine. Potential benefits include accurate definition of the anatomy, demonstration of resection margins, and intraoperative guidance. The technique may prove particularly useful in cases of complex or disrupted anatomy. Copyright 1999 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  1999        PMID: 10631375     DOI: 10.1002/(SICI)1097-0150(1999)4:6<322::AID-IGS4>3.0.CO;2-I

Source DB:  PubMed          Journal:  Comput Aided Surg        ISSN: 1092-9088


  4 in total

Review 1.  Computer-aided navigation in neurosurgery.

Authors:  P Grunert; K Darabi; J Espinosa; R Filippi
Journal:  Neurosurg Rev       Date:  2003-05       Impact factor: 3.042

2.  Towards accurate, robust and practical ultrasound-CT registration of vertebrae for image-guided spine surgery.

Authors:  Charles X B Yan; Benoît Goulet; Julie Pelletier; Sean Jy-Shyang Chen; Donatella Tampieri; D Louis Collins
Journal:  Int J Comput Assist Radiol Surg       Date:  2010-10-26       Impact factor: 2.924

3.  Accuracy of biopsy needle navigation using the Medarpa system--computed tomography reality superimposed on the site of intervention.

Authors:  M Fawad Khan; Selami Dogan; Adel Maataoui; Jessen Gurung; Mirko Schiemann; Hanns Ackermann; Stefan Wesarg; Georgios Sakas; Thomas J Vogl
Journal:  Eur Radiol       Date:  2005-03-08       Impact factor: 5.315

4.  Frameless stereotaxy in a transmandibular, circumglossal, retropharyngeal cervical decompression in a Klippel-Feil patient: technical note.

Authors:  Daniel M Sciubba; Ira M Garonzik; Ian Suk; Gary L Gallia; Anthony Tufaro; Jean Paul Wolinsky; Alex Taghva; Ziya L Gokaslan
Journal:  Eur Spine J       Date:  2006-03-28       Impact factor: 3.134

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.