High-quality filtration allows reuse of anesthesia breathing circuits resulting in cost savings and reduced medical waste.

R Daggan1, A Zefeiridis, D Steinberg, G Larijani, I Gratz, M E Goldberg.   

Abstract

STUDY
OBJECTIVES: To determine if the new Filta-Therm filter prevents contamination and allows the reuse of breathing circuit with considerable cost and environmental savings.
DESIGN: Prospective study. PATIENTS: 52 ASA physical status I, II, III, and IV patients, aged 18 to 75 years.
INTERVENTIONS: Each morning a new breathing circuit was assembled. The Filta-Therm filter (Intersurgical, Inc., Liverpool, NY) elbow, and mask, but not the circuit, were changed between patients. The filter was placed between the Y-piece and the elbow of the breathing circuit. Prior to anesthesia, samples were obtained at the Y-piece, and the inspiratory and expiratory ports of breathing circuit. Following anesthesia, samples were obtained at the Murphy eye of endotrachael tube, and at the Y-piece. The samples were incubated, and the results examined at 24 and 48 hours.
MEASUREMENTS AND MAIN RESULTS: Prior to anesthesia, cultures of the Y-piece and the inspired and expired ports samples showed no growth. Following anesthesia, all 52 samples obtained at the endotracheal tube were contaminated with various organisms, while all 52 Y-piece samples showed negative growth.
CONCLUSIONS: The single use of Filta-Therm filter prevents bacterial contamination and allows reuse of breathing circuit at least twice, resulting in significant cost savings ($50,778 per year). Further studies are needed to establish the safety of reusing breathing circuits when appropriate bacterial filters are used.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10624635     DOI: 10.1016/s0952-8180(99)00083-5

Source DB:  PubMed          Journal:  J Clin Anesth        ISSN: 0952-8180            Impact factor:   9.452


  2 in total

1.  Microbiological safety and cost-effectiveness of weekly breathing circuit changes in combination with heat moisture exchange filters: a prospective longitudinal clinical survey.

Authors:  Nils-Olaf Hübner; Georg Daeschlein; Christian Lehmann; Sergei Musatkin; Ute Kohlheim; Andreas Gibb; Ojan Assadian; Hiroyoshi Kobayashi
Journal:  GMS Krankenhhyg Interdiszip       Date:  2011-12-15

2.  The reuse of anesthesia breathing systems: another difference of opinion and practice between the United States and Europe.

Authors:  Christoph B Egger Halbeis; Alex Macario; John G Brock-Utne
Journal:  J Clin Anesth       Date:  2008-03       Impact factor: 9.452