Estimation of body fluid changes during peritoneal dialysis by segmental bioimpedance analysis.

F Zhu1, D Schneditz, A M Kaufman, N W Levin.   

Abstract

UNLABELLED: Estimation of body fluid changes during peritoneal dialysis by segmental bioimpedance analysis.
BACKGROUND: Commonly used bioimpedance analysis (BIA) is insensitive to changes in peritoneal fluid volume. The purpose of this study was to show, to our knowledge for the first time, that a new segmental approach accurately measures extracellular fluid changes during peritoneal dialysis (PD).
METHODS: Fourteen stable PD patients were studied during a standard exchange with fluids of known conductivity. Bioimpedance was continuously measured in the arm, trunk, and leg and from wrist to ankle. Volume changes were calculated using both a newly developed sum of segmental BIA (SBIA) and current wrist-to-ankle BIA (WBIA) and were compared with actual volume changes measured gravimetrically.
RESULTS: When 2.19 +/- 0.48 L were removed from the peritoneal cavity during draining, 95.2 +/- 13.8% of this volume was detected by SBIA compared with only 12.5 +/- 24. 3% detected by WBIA. When 2.11 +/- 0.20 L of fresh dialysate was infused into the peritoneal cavity during filling, 91.1 +/- 19.6% of this volume was detected by SBIA compared with only 8.8 +/- 21.1% detected by WBIA.
CONCLUSION: The good agreement between measured and calculated data using SBIA was due to: (a) improved placement of electrodes, (b) estimation of trunk extracellular volume based on a new algorithm, and (c) consideration of changes in dialysate conductivity. Correct estimation of fluid volume in the trunk is a prerequisite for applications in which direct analysis of fluid changes cannot be performed such as with peritoneal equilibration tests and continuous flow PD.

Entities:  

Mesh:

Year:  2000        PMID: 10620212     DOI: 10.1046/j.1523-1755.2000.00815.x

Source DB:  PubMed          Journal:  Kidney Int        ISSN: 0085-2538            Impact factor:   10.612


  3 in total

1.  A comparison of fluid status determination using bioelectric impedance and the isotope dilution method in hemodialysis and peritoneal dialysis patients.

Authors:  Yanna Dou; Yanan Gong; Afang Li; Peipei Wang; Dongyan Song; Yuyang Wang; Dong Liu; Genyang Cheng; Zhanzheng Zhao
Journal:  Ren Fail       Date:  2022-12       Impact factor: 3.222

2.  Assessment of body composition in dialysis patients by arm bioimpedance compared to MRI and 40K measurements.

Authors:  M Carter; F Zhu; P Kotanko; M Kuhlmann; L Ramirez; S B Heymsfield; G Handelman; N W Levin
Journal:  Blood Purif       Date:  2009-03-09       Impact factor: 2.614

3.  Monitoring of body water composition by the simultaneous use of bioelectrical impedance analysis and Crit-Line(®) during hemodialysis.

Authors:  Naoki Sugano; Keitaro Yokoyama; Naohiko Kato; Yoichiro Hara; Satoshi Endo; Jun Mitome; Taisei Kin; Goro Tokudome; Satoru Kuriyama; Tatsuo Hosoya; Takashi Yokoo
Journal:  Clin Exp Nephrol       Date:  2014-02-12       Impact factor: 2.801