C H Huang1, Y M Lee, J H Lai, J J Liau, C K Cheng.
Abstract
From 1991 to 1996, 953 cases of porous-coated anatomic modular knee prostheses with all-polyethylene patellar components were implanted. Among them, 4 cases had breakage of the patellar component at the peg-button interfaces. One had loosening of the patellar component by cutting out the patellar bony bed. Heavy body weight, weakness of the pegs of the all-polyethylene patellar component, and osteonecrosis of the patella were conceived as the causes of failure.Entities:
Mesh:
Substances:
Year: 1999 PMID: 10614884 DOI: 10.1016/s0883-5403(99)90007-4