Accuracy of ultrasonography and magnetic resonance imaging in detecting failure of breast implants filled with silicone gel.

W H Beekman1, J J Hage, A H Taets van Amerongen, J W Mulder.   

Abstract

We prospectively studied the accuracy of magnetic resonance imaging (MRI) and ultrasonography (US) for preoperative detection of rupture in 35 single-lumen implants filled with silicone gel in 18 patients. The positive predictive value of US for rupture of an implant was 70% and the negative predictive value 64%. Sensitivity and specificity were 44% and 87%, respectively. Accuracy, defined as the total true positive and true negative values divided by the total number of implants studied was 66%. The positive predictive value of MRI was 100% and the negative predictive value 90%. The corresponding sensitivity and specificity were 88% and 100% and the accuracy 94%. MRI offers significantly better diagnostic sensitivity (p = 0.02) and accuracy (p = 0.004), and should be regarded as the "gold standard" in the evaluation of rupture of breast implants filled with silicone gel. When MRI is not readily available, US is an acceptable alternative.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10614750     DOI: 10.1080/02844319950159127

Source DB:  PubMed          Journal:  Scand J Plast Reconstr Surg Hand Surg        ISSN: 0284-4311


  1 in total

Review 1.  The effect of study design biases on the diagnostic accuracy of magnetic resonance imaging for detecting silicone breast implant ruptures: a meta-analysis.

Authors:  Jae W Song; Hyungjin Myra Kim; Lillian T Bellfi; Kevin C Chung
Journal:  Plast Reconstr Surg       Date:  2011-03       Impact factor: 4.730

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.