Literature DB >> 10612904

A transconductance driven-right-leg circuit.

E M Spinelli1, N H Martínez, M A Mayosky.   

Abstract

Biopotential measurements are very sensitive to electromagnetic interference (EMI). EMI gets into the acquisition system by many ways, both as differential and common mode signals, driven-right-leg circuits (DRL) are widely used to reduce common mode interference. This paper reports an improvement on the classic DRL. The proposed circuit uses a transconductance amplifier to drive the patient's body. This configuration has some interesting properties, which provide an extended bandwidth for high-frequency EMI rejection (such as fluorescent lights interference). The improvement is around 20 dB for frequencies of few kilohertz and the circuit is easy to compensate for stability. A comparative analysis against a typical DRL is presented, the results obtained have been experimentally tested.

Entities:  

Mesh:

Year:  1999        PMID: 10612904     DOI: 10.1109/10.804574

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  3 in total

1.  Estimation of stray coupling capacitances in biopotential measurements.

Authors:  Marcelo Haberman; Augusto Cassino; Enrique Spinelli
Journal:  Med Biol Eng Comput       Date:  2011-07-20       Impact factor: 2.602

2.  Performance of human body communication-based wearable ECG with capacitive coupling electrodes.

Authors:  Jun Sakuma; Daisuke Anzai; Jianqing Wang
Journal:  Healthc Technol Lett       Date:  2016-07-05

3.  Heart Rate Variability Monitoring during Sleep Based on Capacitively Coupled Textile Electrodes on a Bed.

Authors:  Hong Ji Lee; Su Hwan Hwang; Hee Nam Yoon; Won Kyu Lee; Kwang Suk Park
Journal:  Sensors (Basel)       Date:  2015-05-14       Impact factor: 3.576

  3 in total