J M Pouilles1, P Collard, F Tremollieres, P Frayssinet, J J Railhac, J P Cahuzac, A Autefage, C Ribot.
Abstract
We evaluated the precision and accuracy of in vivo measurements of spine bone mineral density (BMD) and bone mineral content (BMC) in five ewes using dual-energy X-ray absorptiometry (DXA, Lunar DPX-L). The short-term in vivo reproducibility expressed as the coefficient of variation (CV) varied from 0.9 to 1.6% for spine BMD and from 1 to 3.1% for spine BMC. The ex vivo measurements, performed in 20 cm of water to simulate soft tissue thickness, correlated closely with the in vivo measurements, yielding an r value of 0.98 and 0.97 for spine BMD and BMC, respectively. The accuracy was determined by comparing the total BMC of each vertebra measured in vivo with the corresponding ash weight. The correlation coefficient between the two measurements was r = 0.98, with an accuracy error of 5.6%. We concluded that the DXA allows a precise and accurate measurement of spine bone mineral in live ewes using the methodology designed for humans.Entities:
Mesh:
Year: 2000 PMID: 10602849 DOI: 10.1007/s002230050014