Central venous catheter placement in children: evaluation of electrocardiography using J-wire.

L Simon1, A Teboul, N Gwinner, G Boulay, S Cerceau-Delaporte, J Hamza.   

Abstract

Misplacement of a central venous catheter may lead to myocardial perforation and dysrhythmia. Atrial electrocardiography (ECG) through a saline column is an effective but complex method to determine the accurate location of catheters. We evaluated a simplified variant of this technique using the guidewire as an internal electrode in 23 children (5-16 years old) undergoing spinal surgery. Catheters were placed using a Seldinger technique after jugular or subclavian venous puncture. Each time the operator recognized the atrial signal, the catheter was found to be correctly placed on the chest radiograph (20/23). In three patients, the atrial signal was not obtained. A technical error was responsible in one case whereas the two others were related to aberrant migrations of the catheter either into a subclavian vein or into the pleura. In this latter case, the complication was unrecognized on the first radiograph despite malposition having been predicted by atrial ECG. We conclude that a method using atrial ECG guidance is sensitive and specific, and may be an alternative to the classical chest radiograph to detect accurate placement of central venous catheters in children.

Entities:  

Mesh:

Year:  1999        PMID: 10597553     DOI: 10.1046/j.1460-9592.1999.00422.x

Source DB:  PubMed          Journal:  Paediatr Anaesth        ISSN: 1155-5645            Impact factor:   2.556


  2 in total

1.  Evaluating ECG-aided tip localization of peripherally inserted central catheter in patients with cancer.

Authors:  Yan-Jin Liu; Lei Dong; Xiao-Ping Lou; Jin-Hong Miao; Xiu-Xia Li; Xiao-Jing Li; Jing Li; Qian-Qian Liu; Zhi-Wei Chang
Journal:  Int J Clin Exp Med       Date:  2015-08-15

Review 2.  Clinical review: vascular access for fluid infusion in children.

Authors:  Nikolaus A Haas
Journal:  Crit Care       Date:  2004-06-03       Impact factor: 9.097

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.