R Marsilio1, R Dall'Amico, G Giordano, L Murer, G Montini, M Ros, L Bacelle, M Plebani, N Dussini, G Zacchello.
Abstract
We report a simple and reliable high performance liquid chromatography method for measuring creatinine in serum and urine. The chromatographic run is performed on a C(18) column after protein precipitation with acetone and addition of cimetidine as an internal standard. The separation is carried out in 20 min at a flow rate of 0.8 ml/min, with a mobile phase consisting of 100 mmol/l sodium dihydrogen phosphate solution, containing 30 mmol/l sodium lauryl sulfate pH 3.0 and acetonitrile (60:36, v/v). The absorbance is monitored at 200 nm. The relationship between creatinine concentration and the creatinine/internal standard peak area is linear up to 1,088 micromol/l. Within-run precision measured at three different creatinine concentrations ranges from 0.89% to 2.34% in serum and from 0.34% to 1.10% in urine. Between-run precision varies from 1.68% to 3.17% in serum and from 1.58% to 1.85% in urine over a wide range of concentrations. Analytical recovery is between 98.71% and 101.25% in serum and between 98.96% and 100.27% in urine. The detection limit is 3.24 micromol/l for a signal-to-noise ratio of 3. The method shows a good linearity with the reference isotope dilution gas chromatography-mass spectrometry procedure (r=0.999), without interferences, even in the presence of high bilirubin concentrations.Entities:
Mesh:
Substances:
Year: 1999 PMID: 10592106 DOI: 10.1007/s005990050072
Source DB: PubMed Journal: Int J Clin Lab Res ISSN: 0940-5437