Literature DB >> 10590003

Laser in situ keratomileusis for myopic astigmatism.

K M Rashad1.   

Abstract

PURPOSE: To evaluate the refractive results of laser in situ keratomileusis (LASIK) for myopic astigmatic eyes, and to assess the efficacy, accuracy, stability, and safety of the procedure.
METHODS: LASIK was performed on 113 eyes of 73 patients for correction of myopic astigmatism ranging from 1.00 to 5.00 D, as measured by manifest refraction, with a mean baseline refractive astigmatism of 2.09 +/- 1.12 D. The Chiron Automated Corneal Shaper was used to create a corneal flap, and laser ablation was performed using the Chiron-Technolas Keracor 116 excimer laser. Follow-up time was 12 months for all eyes.
RESULTS: Refractive astigmatism was stable by 3 months after surgery. At 1 year after LASIK, refractive astigmatism was reduced to a mean of 0.25 +/- 0.31 D (range 0 to 1.00 D). Sixty-one eyes (54%) had no residual astigmatism and 98 eyes (86.7%) had 0 to 0.50 D of refractive astigmatism. The mean percent reduction of preoperative astigmatism was 87.9 +/- 14.9%. The mean axis deviation of the surgically induced astigmatism was 2.1 +/- 3.1 degrees, with 96 eyes (84.9%) within 5 degrees of the desired axis. The percent correction of preoperative astigmatism in the proper axis was 97.1 +/- 15.5%. Spectacle-corrected visual acuity improved by 2 lines in 11 eyes (9.7%), and was reduced by 1 line only in 1 eye. There were no other significant complications.
CONCLUSION: LASIK with the Chiron-Technolas Keracor 116 excimer laser was effective for correction of myopic astigmatism, with good stability after 3 months. The results were predictable with an acceptable degree of accuracy. LASIK is a safe procedure with very few complications.

Entities:  

Mesh:

Year:  1999        PMID: 10590003     DOI: 10.3928/1081-597X-19991101-09

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  4 in total

1.  Opposite Clear Corneal Incisions versus Steep Meridian Incision Phacoemulsification for Correction of Pre-existing Astigmatism.

Authors:  Noushin Bazzazi; Behzad Barazandeh; Mani Kashani; Maryam Rasouli
Journal:  J Ophthalmic Vis Res       Date:  2008-04

2.  Increased preference of surface ablation over laser in situ keratomileusis between 2008-2011 is correlated to risk of ecatasia.

Authors:  Elad Moisseiev; Tzahi Sela; Liza Minkev; David Varssano
Journal:  Clin Ophthalmol       Date:  2013-01-10

3.  Evaluation of the prophylactic use of mitomycin-C to inhibit haze formation after photorefractive keratectomy in high myopia: a prospective clinical study.

Authors:  Hassan Hashemi; Seied Mohammad Reza Taheri; Akbar Fotouhi; Azita Kheiltash
Journal:  BMC Ophthalmol       Date:  2004-09-14       Impact factor: 2.209

4.  Posterior corneal astigmatism: a review article.

Authors:  Seyed-Farzad Mohammadi; Masoud Khorrami-Nejad; Moein Hamidirad
Journal:  Clin Optom (Auckl)       Date:  2019-08-12
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.