Tubing failure during prolonged roller pump use: a laboratory study.

G J Peek1, K Wong, C Morrison, H M Killer, R K Firmin.   

Abstract

Little is known about the mechanical forces acting on extracorporeal circuit tubing with prolonged roller pump use during extracorporeal membrane oxygenation (ECMO). We examined the time to tubing rupture of three different materials during actual roller pump use, mean and standard deviation (SD) (SD shown in parentheses): Tygon (control) 243.7 h (175.4); LVA 121 h (14.3); and SRT 6.6 h (2.1). Failure times for both LVA and SRT were significantly different from the control (paired t-test, p = 0.02 and p < 0.001, respectively). The minimum failure times for Tygon and LVA were 99 and 101 h, respectively. We then examined Tygon under conditions of pure compression, demonstrating that even after 3.67 million compression cycles at full occlusion crack formation did not occur. If the tubing was over-occluded, cracks appeared within 24 h. Scanning electron microscopy of Tygon, which has been used during clinical ECMO, and the failure pattern during destruction testing demonstrate that shear stress and compression coexist during clinical ECMO. Use of under-occlusive pump settings could improve tubing life.

Entities:  

Mesh:

Year:  1999        PMID: 10585152     DOI: 10.1177/026765919901400607

Source DB:  PubMed          Journal:  Perfusion        ISSN: 0267-6591            Impact factor:   1.972


  5 in total

1.  A Compression Valve for Sanitary Control of Fluid-Driven Actuators.

Authors:  Simone Calò; James H Chandler; Federico Campisano; Keith L Obstein; Pietro Valdastri
Journal:  IEEE ASME Trans Mechatron       Date:  2019-12-17       Impact factor: 5.303

2.  Rupture of extra-corporeal circuit tubing during cardiopulmonary bypass.

Authors:  Cheemalapati Sai Krishna; Palli Venkata Naresh Kumar; Soumya Kanta Satpathy; Kanteti Ram Mohan; Vedangi Ramesh Babu
Journal:  J Extra Corpor Technol       Date:  2008-03

3.  Using a roller pump for establishing extra-corporal membrane oxygenation (ECMO) - technical considerations for times of crisis.

Authors:  Mirko Kaluza; Benjamin May; Torsten Doenst
Journal:  Perfusion       Date:  2021-02-26       Impact factor: 1.581

4.  Use of roller pump in venovenous extracorporeal membrane oxygenation as an emergency rescue procedure.

Authors:  Surendra Patel; Madhusudan Katti; Pramod Chandolia; Kamlesh Panwar; Alok Kumar Sharma
Journal:  Indian J Thorac Cardiovasc Surg       Date:  2022-08-31

5.  A novel roller pump for physiological flow.

Authors:  Albert Chong; Zhonghua Sun; Lennart van de Velde; Shirley Jansen; Michel Versluis; Michel M P J Reijnen; Erik Groot Jebbink
Journal:  Artif Organs       Date:  2020-03-25       Impact factor: 3.094

  5 in total