Literature DB >> 10575722

[Comparative biomechanical study of 3 types of osteosynthesis of the Duparc grade IV fractures of the calcaneus: value of triangular internal fixation].

R Badet1, F Ribeiro, C Rumelhart, Y Tourne, A Badulescu, D Saragaglia.   

Abstract

PURPOSE OF THE STUDY: We present an in vitro biomechanical study performed to evaluate and compare, for an experimentally produced fracture of the calcaneum (Duparc grade IV), the reaction of 3 standard models of internal fixation commonly used in these fractures and which occupy different volumes.
MATERIALS AND METHODS: We compared different methods of fixation using fresh human calcanei. In two experimental series, we compared triangular internal fixation (3 1/4 tube AO plates Saragaglia), Y internal fixation (2 1/3 tube AO plates Bezes), isolated screw technique (three 3.5 diameter screws, two 4.5 diameter screws). The plates and screws were made of identical material (316L). Both series used 8 pairs of bone (talo-calcaneum system) with the same fracture submitted to a 200N to 1000N load. Stiffness and movement were analyzed using 8 references on the calcaneum.
RESULTS: The stiffness and movement analysis with 8 references points demonstrated the superior resistance to bending with the triangular internal fixation. Fixation stability was significantly better than with the Y or screw technique. DISCUSSION: This study underlined the importance of triangular trabecular organization of cancellous bone on calcaneum biomechanics. We showed that the 3 (anterior, posterior and inferior) trabeculae must be repaired in calcaneum fractures to achieve horizontal and vertical stability of the talar joint.
CONCLUSION: In our hands, restoration of the triangular architecture of calcaneum fractures, to resemble a roof truss, where the talus is fixed to the triangle vertex, is fundamental to obtain a rigid and stable internal fixation.

Entities:  

Mesh:

Year:  1999        PMID: 10575722

Source DB:  PubMed          Journal:  Rev Chir Orthop Reparatrice Appar Mot        ISSN: 0035-1040


  2 in total

1.  A biomechanical comparison of conventional versus an anatomic plate and compression bolts for fixation of intra-articular calcaneal fractures.

Authors:  Haili Wang; Zhaoxu Yang; Zhanpo Wu; Wei Chen; Qi Zhang; Ming Li; Zhiyong Li; Yingze Zhang
Journal:  J Huazhong Univ Sci Technolog Med Sci       Date:  2012-08-11

2.  [Surgical treatment of articular fractures of the calcaneus by bone plate].

Authors:  Nassreddine Hammou; Hatim Abid; Mohammed Shimi; Abdelhalim El Ibrahimi; Abdelmajid El Mrini
Journal:  Pan Afr Med J       Date:  2015-03-25
  2 in total