Literature DB >> 10569161

Ocular biocompatibility testing of intraocular lenses: a 1 year study in pseudophakic rabbit eyes.

J N Norton1, T Kohnen, R B Hackett, A Patel, D D Koch.   

Abstract

PURPOSE: To evaluate the appropriate duration for conducting ocular biocompatibility studies with an intraocular lens (IOL) in the pseudophakic rabbit model.
SETTING: Alcon Laboratories, Inc., Fort Worth, Texas, USA.
METHODS: A single-piece biconvex poly(methyl methacrylate) (PMMA) IOL was implanted in the capsular bag of 18 eyes of New Zealand white rabbits; 8 eyes received sham surgeries. Rabbits were monitored clinically and then sacrificed 6 or 12 months after surgery for histopathological examination of ocular tissues.
RESULTS: Biomicroscopic examination revealed mild ocular changes in all surgical eyes during the first 3 months postoperatively. After that, there was a high incidence of posterior synechias, flare, and posterior capsule opacification (PCO) in eyes with PMMA IOLs. Posterior synechias and flare scores remained mild to moderate throughout the study, whereas PCO severity increased over time. Similar findings were observed in sham eyes. In addition, several eyes with PMMA IOLs developed IOL dislocation, hyphema, iris bombe, and a fibrous membrane covering the IOL. No discernible differences in biomicroscopic scores were observed in eyes at 6 or 12 months. Intraocular pressures and morphology of the corneal endothelium were normal in both groups. Results from histopathological analysis of the ocular tissues were consistent with observations from the biomicroscopic examinations.
CONCLUSION: The results suggest that IOL implantation studies in the pseudophakic rabbit eye should be conducted for 3 or fewer months. Regulatory guidelines requiring longer ocular implantation studies should be revised to reflect the inherently rapid and extensive ocular response in the rabbit model.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10569161     DOI: 10.1016/s0886-3350(99)00237-0

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  A 1-year study on carbon, titanium surface-modified intraocular lens in rabbit eyes.

Authors:  Zhaoxu Yuan; Huimin Sun; Jiaqin Yuan
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2004-09-10       Impact factor: 3.117

2.  A novel suprachoroidal microinvasive glaucoma implant: in vivo biocompatibility and biointegration.

Authors:  Ian Grierson; Don Minckler; Marian K Rippy; Andrew J Marshall; Nathalie Collignon; Jessica Bianco; Benoit Detry; Murray A Johnstone
Journal:  BMC Biomed Eng       Date:  2020-10-14

Review 3.  Surface Modification of Intraocular Lenses.

Authors:  Qi Huang; George Pak-Man Cheng; Kin Chiu; Gui-Qin Wang
Journal:  Chin Med J (Engl)       Date:  2016-01-20       Impact factor: 2.628

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.