Potential for chlorate interference in ion chromatographic determination of total nitrogen in natural waters following alkaline persulfate digestion.

J A Halstead1, J Edwards, R J Soracco, R W Armstrong.   

Abstract

Determination of total nitrogen in aqueous samples after thermal potassium peroxydisulfate (persulfate) digestion is a commonly used alternative to the tedious Kjeldahl procedure. When ion chromatography is used to quantify the nitrate formed during digestion, there is a potential for interference from a chlorate peak if the digested sample initially contained chloride in concentrations close to or greater than the concentration of nitrogen. It was determined that this interference can be avoided either by using chromatographic conditions which cleanly resolve the nitrate and chlorate peaks (e.g., the Dionex AG9-HG column) or by using digestion reagent concentrations chosen to maintain a high pH throughout the digestion. The second alternative is not a viable option for investigators using a single digestion for both total nitrogen (TN) and total phosphorus (TP) analysis.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10536854     DOI: 10.1016/s0021-9673(99)00757-8