Treatment head design for multileaf collimated high-energy electrons.

M G Karlsson1, M Karlsson, C M Ma.   

Abstract

This paper describes how a conventional treatment head can be modified for use of multileaf collimated electron beams. Automatic and dynamic beam delivery are possible for both electrons and photons by using the computer controlled multileaf collimator (MLC) for both photon and electron beams. Thereby, the electron beams can be mixed more freely into the treatment to take advantage of the specific depth modulation characteristics of electrons. The investigation was based on Monte Carlo calculations using the software package BEAM. The physical parameters used in this optimization were the beam penumbra and the virtual/effective point source position. These parameters are essential for shaping beams, beam matching and for dosimetry calculations. The optimization was carried out by modifying a number of parameters: replacing the air atmosphere in the treatment head with helium, adding a helium bag below the MLC, changing the position of the scattering foils, modifying the monitor chamber, and adjusting the position of the MLC. The beam characteristics for some of these designs were found to fulfil our criteria for clinically useful beams down to at least 9 MeV.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10535633     DOI: 10.1118/1.598732

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Monte Carlo simulations of electron beams collimated with a dual electron multileaf collimator: a feasibility study.

Authors:  S O Inyang; A C Chamberlain
Journal:  Radiol Phys Technol       Date:  2009-07-04

2.  Dosimetric characteristics of an electron multileaf collimator for modulated electron radiation therapy.

Authors:  Ahmed Abdel Rahman Eldib; Mohamed I ElGohary; Jiajin Fan; Lihui Jin; Jinsheng Li; Charlie Ma; Nader A Elsherbini
Journal:  J Appl Clin Med Phys       Date:  2010-04-12       Impact factor: 2.102

3.  Measurement and Monte Carlo simulation for energy- and intensity-modulated electron radiotherapy delivered by a computer-controlled electron multileaf collimator.

Authors:  Lihui Jin; Ahmed Eldib; Jinsheng Li; Ismail Emam; Jiajin Fan; Lu Wang; C-M Ma
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

4.  Design and production of 3D printed bolus for electron radiation therapy.

Authors:  Shiqin Su; Kathryn Moran; James L Robar
Journal:  J Appl Clin Med Phys       Date:  2014-07-08       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.