In vitro and in vivo evaluation of a new polysulfone membrane for hemodialysis. Reference methodology and clinical results. (Part 1: in vitro study).

C Ronco1, A Brendolan, G Cappelli, M Ballestri, P Inguaggiato, L Fortunato, M Milan, G Pietribiasi, G La Greca.   

Abstract

Different high flux membranes have been recently developed. The present study is aimed at describing the technical features and the clinical performances of a new high flux polysulfone membrane (T-sulfone, Toray, Japan). The study has been carried out on two different dialyzers (surface area = 1.3 and 1.8 m2). The filters have been tested in vitro under definite experimental conditions. The hydraulic flow resistance, the pressure drop in the blood compartment and the hydraulic permeability have been determined in a wide range of in vitro experimental conditions. The in vitro sieving coefficients for various solutes have also been determined utilizing human blood. Hydraulic permeability was found in the range of 28.4 ml/h/mm Hg/m2 and sieving coefficients were between 0.96 and 1.0 for all low molecular weight solutes. The sieving coefficient for inulin was 0.95. The pressure drop in the filter at 300 ml/min of blood flow was 95 mm Hg for the 1.3 m2 and 57 mm Hg for the 1.8 m2. The filters are then designed to operate in the presence of high blood flows without excessive resistance in the blood compartment. The blood compartment analyzed by means of a special radiological sequence obtained with a helical scanner after dye injection confirmed the homogeneous distribution of the blood flow in several cross sections of the bundle. Adequate distribution of dialysate was confirmed with a similar method applied to the dialysate compartment. The new imaging techniques utilized were greatly helpful to determine adequacy of filter design and flows distribution.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10532429

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595


  2 in total

1.  Performance of polysulfone membrane dialyzers and dialysate flow pattern.

Authors:  Ayumu Nakashima; Satoshi Ogata; Shigehiro Doi; Michihiro Yamahira; Satoe Naraki; Norihisa Takasugi; Toshifumi Ohmoto; Takafumi Ito; Takao Masaki; Noriaki Yorioka
Journal:  Clin Exp Nephrol       Date:  2006-09       Impact factor: 2.801

Review 2.  [Antibiotic dosing for renal function disorders and continuous renal replacement therapy].

Authors:  Erik Michael; Detlef Kindgen-Milles
Journal:  Anaesthesist       Date:  2015-04       Impact factor: 1.041

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.