Literature DB >> 10531164

Self-adjusting nasal continuous positive airway pressure therapy based on measurement of impedance: A comparison of two different maximum pressure levels.

W J Randerath1, K Parys, F Feldmeyer, B Sanner, K H Rühle.   

Abstract

STUDY
OBJECTIVE: Automatic titration using the forced oscillation technique (FOT) has recently been developed for the treatment of obstructive sleep apnea syndrome (OSAS). So far, it is not known if therapy with automatic nasal continuous positive airway pressure (nCPAP) using a preset upper pressure limitation or a free range (which might lead to higher mean pressure) is preferable with regard to obstructive events, sleep stages, and pressure characteristics.
DESIGN: After diagnostic polysomnography, patients were randomly assigned to two settings with the self-adjusting nCPAP (APAP) device based on the FOT. In mode 1, the pressure variation ranged from 4 to 15.5 cm H(2)O, and in mode 2, the pressure variation ranged from 4 cm H(2)O to an individual upper pressure limit. PATIENTS: Eleven men, aged 53.0 +/- 6.8 years with a body mass index of 32.4 +/- 5.1 kg/m(2) and an apnea-hypopnea index (AHI) of 31.6 +/- 26.6/h. MEASUREMENTS AND
RESULTS: Manually titrated pressure was at 9.3 +/- 2.1 cm H(2)O, the mean pressure in mode 1 was 5.4 +/- 1.0 cm H(2)O (p < 0.01), and the mean pressure in mode 2 was 5.1 +/- 0.7 cm H(2)O (p < 0.01). A reduction of respiratory events (baseline AHI, 31.6 +/- 26.6/h; AHI in mode 1, 3.4 +/- 4.5; AHI in mode 2, 5.0 +/- 7.2; each with p < 0.001) and an increase in the "rapid eye movement" stage of sleep (baseline, 13.0 +/- 5.5%; mode 1, 22.0 +/- 7.7 [p < 0. 05]; mode 2, 23.0 +/- 7.9 [p < 0.01]) were achieved. In mode 1, the mean pressure was below the manual pressure 91.7 +/- 9.3% of the time, and in mode 2, the mean pressure was below the manual pressure 90.4 +/- 6.3% of the time. The manual pressure was exceeded by 5.5 +/- 7.4% (mode 1) and by 5.2 +/- 3.1% (mode 2).
CONCLUSION: We conclude that nCPAP therapy based on the FOT permits the adequate treatment of OSAS with significantly lower pressure than manually titrated nCPAP therapy does. A presetting of an upper pressure limit has no advantage compared to free range.

Entities:  

Mesh:

Year:  1999        PMID: 10531164     DOI: 10.1378/chest.116.4.991

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  4 in total

Review 1.  Sleep.7: positive airway pressure therapy for obstructive sleep apnoea/hypopnoea syndrome.

Authors:  P Gordon; M H Sanders
Journal:  Thorax       Date:  2005-01       Impact factor: 9.139

Review 2.  Effect of automatic versus fixed continuous positive airway pressure for the treatment of obstructive sleep apnea: an up-to-date meta-analysis.

Authors:  Ting Xu; Taoping Li; Dongning Wei; Yuan Feng; Lewu Xian; Haiqing Wu; Jian Xu
Journal:  Sleep Breath       Date:  2011-12-03       Impact factor: 2.816

3.  New technologies to detect static and dynamic upper airway obstruction during sleep.

Authors:  J M Montserrat; R Farré; D Navajas
Journal:  Sleep Breath       Date:  2001-12       Impact factor: 2.816

4.  Pressure modification or humidification for improving usage of continuous positive airway pressure machines in adults with obstructive sleep apnoea.

Authors:  Barry Kennedy; Toby J Lasserson; Dariusz R Wozniak; Ian Smith
Journal:  Cochrane Database Syst Rev       Date:  2019-12-02
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.