Literature DB >> 10528832

Identifying delayed separation in plasma homocysteine specimens.

M Chen1, K L Nuttall.   

Abstract

The delayed separation of plasma from the cellular components of blood can lead to falsely elevated homocysteine results. The incidence of delayed separation in patient specimens was examined by using the ratio of arginine to the sum of arginine plus ornithine [arg/(arg + orn)], a ratio <0.50 being consistent with delayed separation. Two groups were examined: 1) low homocysteine (<10 micromol/L, n = 10), and 2) elevated (>18 micromol/L, n = 12). Specimens in the low group showed an average ratio of 0.47, with 6 of 10 <0.50. In contrast, specimens in the elevated group showed an average ratio of 0.31, with 11 of 12 <0.50. Characteristics consistent with delayed separation occurred frequently, but were significantly more frequent (p = 0.009) in the elevated group. This suggests that many homocysteine results may be falsely elevated due to preanalytic collection problems.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10528832

Source DB:  PubMed          Journal:  Ann Clin Lab Sci        ISSN: 0091-7370            Impact factor:   1.256


  1 in total

1.  Homocysteine and masked hypertension.

Authors:  Beuy Joob; Viroj Wiwanitkit
Journal:  Anatol J Cardiol       Date:  2014-12-25       Impact factor: 1.596

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.