Literature DB >> 10517553

Detection of lysergic acid diethylamide (LSD) in urine by gas chromatography-ion trap tandem mass spectrometry.

J H Sklerov1, K S Kalasinsky, C A Ehorn.   

Abstract

A confirmatory method for the detection and quantitation of lysergic acid diethylamide (LSD) is presented. The method employs gas chromatography-tandem mass spectrometry (GC-MS-MS) using an internal ionization ion trap detector for sensitive MS-MS-in-time measurements of LSD extracted from urine. Following a single-step solid-phase extraction of 5 mL of urine, underivatized LSD can be measured with limits of quantitation and detection of 80 and 20 pg/mL, respectively. Temperature-programmed on-column injections of urine extracts were linear over the concentration range 20-2000 pg/mL (r2 = 0.999). Intraday and interday coefficients of variation were < 6% and < 13%, respectively. This procedure has been applied to quality-control specimens and LSD-positive samples in this laboratory. Comparisons with alternate GC-MS methods and extraction procedures are discussed.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10517553     DOI: 10.1093/jat/23.6.474

Source DB:  PubMed          Journal:  J Anal Toxicol        ISSN: 0146-4760            Impact factor:   3.367


  1 in total

1.  LC-ESI-MS/MS on an ion trap for the determination of LSD, iso-LSD, nor-LSD and 2-oxo-3-hydroxy-LSD in blood, urine and vitreous humor.

Authors:  Donata Favretto; Giampietro Frison; Sergio Maietti; Santo Davide Ferrara
Journal:  Int J Legal Med       Date:  2006-02-23       Impact factor: 2.791

  1 in total