Literature DB >> 10506392

[Lead-related complications in 340 patients with an implantable cardiverter/defibrillator].

H Schwacke1, A Drögemüller, G Siemon, C Werling, W Saggau, J Senges, K Seidl.   

Abstract

From February 1991 to May 1998, 340 patients had a cardioverter/defibrillator (ICD) implanted. Mean age was 64 +/- 9 years, 278 male and 62 female. 60% of patients had coronary artery disease and 31% dilatative cardiomyopathy. Ejection fraction was 38 +/- 14%. The indication for an implantable cardioverter/defibrillator was in 57% of patients ventricular tachycardia, in 43% ventricular fibrillation. 298 patients had a single-chamber cardioverter/defibrillator implanted, 42 patients a dual-chamber cardioverter/defibrillator. In 25 patients additional subcutaneous patch or array electrodes and in 2 patients additional epicardial patch electrodes were implanted. Implantation site was in 92 patients abdominal and in 248 pectoral. Over a period of 7 years 34 lead-related complications occurred in 33 patients (9.7%), after a median of 2 months after implantation. Diagnosis was made by routine chest x-ray in about 55% of lead-related complications, by clinical presentation (inadequate therapy, pain) in 24%, and by electrical parameters in 21%. In patients with an abdominal implantation site, lead-related complications occurred in 20%, in contrast to 6% in patients with a pectoral implantation site. Regarding patients with pectoral implantation site, lead-related complications were observed in 12% of patients with a dual-chamber ICD vs 4% with a single-chamber ICD (p = 0.05), due to dislocation of atrial electrodes with dual-chamber ICD. There were no differences in clinical parameters between patients with pectoral vs abdominal and between single vs dual-chamber ICD. There were no deaths due to lead-related complications.

Entities:  

Mesh:

Year:  1999        PMID: 10506392     DOI: 10.1007/s003920050326

Source DB:  PubMed          Journal:  Z Kardiol        ISSN: 0300-5860


  3 in total

1.  [Not Available].

Authors:  W Kamke; M Schranz; C Dovifat; H Völler
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2000-01

Review 2.  [ICD lead defects: diagnosis and therapeutical options].

Authors:  Christian G Wollmann; Dirk Böcker; Andreas Löher; Hans H Scheld; Günter Breithardt; Rainer Gradaus
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2009-02-11

Review 3.  Algorithms for better arrhythmia discrimination in implantable cardioverter defibrillators.

Authors:  R Schimpf; C Wolpert; B Lüderitz
Journal:  Curr Cardiol Rep       Date:  2001-11       Impact factor: 2.931

  3 in total