Accuracy of intravascular microcatheter pressure measurements: an experimental study.

H Henkes1, S R Felber, K U Wentz, F Czerwinski, H Monstadt, W Weber, D Kühne.   

Abstract

Intravascular pressure measurements are considered useful for the monitoring and assessment of endovascular treatment effects in intracranial vascular malformations. Experimental data on the accuracy of these measurements are limited. A flow phantom with defined intraluminal pressures and pulsatility flow waveforms was used in this study. Microcatheters commercially available for neuroendovascular procedures (length 140-155 cm), with different outer (0.5-0.83 mm) and inner (0.3-0.53 mm) diameters, were introduced into the phantom in the direction of flow. In a static experiment, pressure values from 0 to 75 mmHg were applied, and in the dynamic part of the experiment mean pressure values from 25 to 65 mmHg, with a pulsatile amplitude from 70 to 170 mmHg were employed. In the static experiment, there was a linear relationship between the pressure values obtained through the microcatheters and the local transducer of the flow phantom. The pulsatile experiments showed increased damping of the pressure waveforms with decreasing inner diameter of the microcatheters. However, the mean pressure values remained accurate. This experimental study has shown that mean pressure values can be accurately measured through microcatheters from 0.3-0.5 mm inner diameter and more than 140 cm in length. In vivo pressure measurements during interventional procedures are therefore reproducible and can be used for monitoring of embolization effects in patients.

Entities:  

Mesh:

Year:  1999        PMID: 10505007     DOI: 10.1259/bjr.72.857.10505007

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  4 in total

1.  Intravascular pressure measurements in feeding pedicles of brain arteriovenous malformations.

Authors:  H Henkes; T F Gotwald; S Brew; E Miloslavski; F Kämmerer; D Kühne
Journal:  Neuroradiology       Date:  2005-11-22       Impact factor: 2.804

2.  Pressure measurements in arterial feeders of brain arteriovenous malformations before and after endovascular embolization.

Authors:  H Henkes; T F Gotwald; S Brew; F Kaemmerer; E Miloslavski; D Kuehne
Journal:  Neuroradiology       Date:  2004-06-15       Impact factor: 2.804

3.  Intra-arterial pulse wave analysis during thrombectomy for the assessment of collateral status - A feasibility study.

Authors:  Johanna Sandmann; Thorsten Sichtermann; Franziska Sonja Müschenich; Hadi Nasri; Sarah Heringer; Anastasios Mpotsaris; Martin Kramer; Martin Wiesmann; Omid Nikoubashman
Journal:  PLoS One       Date:  2019-01-10       Impact factor: 3.240

4.  Digital intravascular pressure wave recording during endovascular treatment reveals abnormal shunting flow in vertebral venous fistula of the vertebral artery: illustrative case.

Authors:  Yoshiteru Shimoda; Shinya Sonobe; Kuniyasu Niizuma; Toshiki Endo; Hidenori Endo; Mayuko Otomo; Teiji Tominaga
Journal:  J Neurosurg Case Lessons       Date:  2021-07-12
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.