Scatter fractions from linear accelerators with x-ray energies from 6 to 24 MV.

P L Taylor1, J E Rodgers, J Shobe.   

Abstract

Computation of shielding requirements for a linear accelerator must take into account the amount of radiation scattered from the patient to areas outside the primary beam. Currently, the most frequently used data are from NCRP 49 that only includes data for x-ray energies up to 6 MV and angles from 30 degrees to 135 degrees. In this work we have determined by Monte Carlo simulation the scattered fractions of dose for a wide range of energies and angles of clinical significance including 6, 10, 18, and 24 MV and scattering angles from 10 degrees to 150 degrees. Calculations were made for a 400 cm2 circular field size impinging onto a spherical phantom. Scattered fractions of dose were determined at 1 m from the phantom. Angles from 10 degrees to 30 degrees are of concern for higher energies where the scatter is primarily in the forward direction. An error in scatter fraction may result in too little secondary shielding near the junction with the primary barrier. The Monte Carlo code ITS (Version 3.0) developed at Sandia National Laboratory and NIST was used to simulate scatter from the patient to the barrier. Of significance was the variation of calculated scattered dose with depth of measurement within the barrier indicating that accurate values may be difficult to obtain. Mean energies of scatter x-ray spectra are presented.

Entities:  

Mesh:

Year:  1999        PMID: 10501042     DOI: 10.1118/1.598670

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Concurrent kilovoltage CBCT imaging and megavoltage beam delivery: suppression of cross-scatter with 2D antiscatter grids and grid-based scatter sampling.

Authors:  Farhang Bayat; Mohamed Elsayed Eldib; Brian Kavanagh; Moyed Miften; Cem Altunbas
Journal:  Phys Med Biol       Date:  2022-08-09       Impact factor: 4.174

2.  Megavoltage cross-scatter rejection and correction using 2D antiscatter grids in kilovoltage CBCT imaging.

Authors:  Farhang Bayat; Mohamed Elsayed Eldib; Cem Altunbas
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2022-04-04

3.  Radiation therapy vault shielding calculational methods when IMRT and TBI procedures contribute.

Authors:  J E Rodgers
Journal:  J Appl Clin Med Phys       Date:  2001       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.