A diamond detector in the dosimetry of high-energy electron and photon beams.

W U Laub1, T W Kaulich, F Nüsslin.   

Abstract

A diamond detector type 60003 (PTW Freiburg) was examined for the purpose of dosimetry with 4-20 MeV electron beams and 4-25 MV photon beams. Results were compared with those obtained by using a Markus chamber for electron beams and an ionization chamber for photon beams. Dose distributions were measured in a water phantom with the detector connected to a Unidos electrometer (PTW Freiburg). After a pre-irradiation of about 5 Gy the diamond detector shows a stability in response which is better than that of an ionization chamber. The current of the diamond detector was measured under variation of photon beam dose rate between 0.1 and 7 Gy min(-1). Different FSDs were chosen. Furthermore the pulse repetition frequency and the depth of the detector were changed. The electron beam dose rate was varied between 0.23 and 4.6 Gy min(-1) by changing the pulse-repetition frequency. The response shows no energy dependence within the covered photon-beam energy range. Between 4 MeV and 18 MeV electron beam energy it shows only a small energy dependence of about 2%, as expected from theory. For smaller electron energies the response increases significantly and an influence of the contact material used for the diamond detector can be surmised. A slight sublinearity of the current and dose rate was found. Detector current and dose rate are related by the expression i alpha Ddelta, where i is the detector current, D is the dose rate and delta is a correction factor of approximately 0.963. Depth-dose curves of photon beams, measured with the diamond detector, show a slight overestimation compared with measurements with the ionization chamber. This overestimation is compensated for by the above correction term. The superior spatial resolution of the diamond detector leads to minor deviations between depth-dose curves of electron beams measured with a Markus chamber and a diamond detector.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10495113     DOI: 10.1088/0031-9155/44/9/306

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  4 in total

1.  A simple and fast physics-based analytical method to calculate therapeutic and stray doses from external beam, megavoltage x-ray therapy.

Authors:  Lydia J Jagetic; Wayne D Newhauser
Journal:  Phys Med Biol       Date:  2015-06-04       Impact factor: 3.609

2.  Silicon diodes as an alternative to diamond detectors for depth dose curves and profile measurements of photon and electron radiation.

Authors:  Christian Scherf; Christiane Peter; Jussi Moog; Jörg Licher; Eugen Kara; Klemens Zink; Claus Rödel; Ulla Ramm
Journal:  Strahlenther Onkol       Date:  2009-08-04       Impact factor: 3.621

Review 3.  From cultured to uncultured genome sequences: metagenomics and modeling microbial ecosystems.

Authors:  Daniel R Garza; Bas E Dutilh
Journal:  Cell Mol Life Sci       Date:  2015-08-09       Impact factor: 9.261

4.  Different Dosimeters/Detectors Used in Small-Field Dosimetry: Pros and Cons.

Authors:  Wrya Parwaie; Soheila Refahi; Mahdieh Afkhami Ardekani; Bagher Farhood
Journal:  J Med Signals Sens       Date:  2018 Jul-Sep
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.