Comparison of two inspiratory: expiratory ratios during high frequency oscillation.

G Dimitriou1, A Greenough, V Kavvadia, A D Milner.   

Abstract

UNLABELLED: The aim of this study was to compare gas exchange and volume delivery during high frequency oscillation at two frequently used inspiratory:expiratory (I:E) ratios: 1:2 and 1:1, other oscillatory settings being kept constant. A group of 13 infants with respiratory distress syndrome, median gestational age 28 weeks (range 23-36) and postnatal age 1 day (range 1-8) were studied. At the I:E ratio of 1:1 compared to 1:2 the median paCO(2) was lower, P < 0.05 (30 mmHg, range 22-47 vs 34 mmHg, range 27-46) and the volume delivered higher, P < 0.01 (2.6 ml/kg, range 1.2-5.6 vs 2.0 ml/kg, range 1.0-3.9). There was no significant difference in oxygenation levels at the two I:E ratios. In a related in vitro study, changing the I:E ratio from 1:2 to 1:1 increased the mean airway pressure by a median of 8.6% (range 2.9-28.1%).
CONCLUSION: Routinely maintained longer expiratory than inspiratory times during high frequency oscillation should be discouraged.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10486079     DOI: 10.1007/s004310051208

Source DB:  PubMed          Journal:  Eur J Pediatr        ISSN: 0340-6199            Impact factor:   3.183


  2 in total

Review 1.  Update on modalities of mechanical ventilators.

Authors:  A Greenough
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2002-07       Impact factor: 5.747

Review 2.  High-frequency oscillatory ventilation: what large-animal studies have taught us!

Authors:  Robert M Kacmarek; Atul Malhotra
Journal:  Crit Care Med       Date:  2005-03       Impact factor: 7.598

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.