Clinical validation of an automatic device measuring blood pressure in the fingers.

E Arosio1, M Zannoni.   

Abstract

BACKGROUND: To assess the agreement between a new automatic device (FS-20D) using a cuff-oscillometric method to measure arterial blood pressure (BP) in the fingers and a standard mercury sphygmomanometer.
METHODS: The blood pressure measurements were taken in a sequential order, in a sample of both normotensive subjects (n. 57) and slight to moderate hypertensive patients (n. 28) without vascular complications.
RESULTS: The mean sphygmomanometer-monitor difference was 0.52 +/- 4.57 mmHg for systolic and 0.25 +/- 4.41 mmHg for diastolic values; the agreement limits were: SBP -8.6 divided by 9.6 mmHg, 95% CI: -0.5 divided by 1.5; DBP: -8.6 divided by 9.1 mmHg, 95% CI: -0.7 divided by 1.2. The grade of agreement between the monitor and the sphygmomanometer was "A" (British Hypertension Society) for both systolic and diastolic values (difference of readings < 5 mmHg: 82%; < 10 mmHg: 97% for systolic blood pressure, 98% of diastolic blood pressure).
CONCLUSIONS: The monitor was proved to be reliable with a good level of precision and accuracy. The FS-20D monitor may be used in self-monitoring of blood pressure of patients with slight to moderate hypertension.

Entities:  

Mesh:

Year:  1999        PMID: 10479914

Source DB:  PubMed          Journal:  Panminerva Med        ISSN: 0031-0808            Impact factor:   5.197


  1 in total

1.  Continuous mean arterial pressure measurement in the fingers: the influence of local arm cooling.

Authors:  R Raamat; K Jagomägi; J Talts; E Länsimies; J Jurvelin; P Kolari
Journal:  Med Biol Eng Comput       Date:  2001-09       Impact factor: 2.602

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.