Implant failure of the gamma nail.

C Gaebler1, S Stanzl-Tschegg, E K Tschegg, C Kukla, W A Menth-Chiari, G E Wozasek, T Heinz.   

Abstract

The gamma nail is a temporary implant characterised by a limited life expectancy under continuous dynamic stress. We reviewed a series of 839 patients with gamma nail stabilisation and found two fatigue fractures (0.2%) at the aperture of the distal locking holes. This complication has not been described in the literature. Metallurgic and scanning electron microscopic examinations proved that the fatigue zones occurred at the clover-leaf grooves, which is where the diameter of the gamma nail is reduced. The clover-leaf diameter is of no biomechanical use in gamma nail stabilisation. We suggested product modification of the gamma nail to produce implants with a round diameter instead of a clover-leaf shape. A modified implant is already in use at our institution.

Entities:  

Mesh:

Year:  1999        PMID: 10476276     DOI: 10.1016/s0020-1383(98)00215-0

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  12 in total

1.  [Partial weight-bearing in rehabilitation. Strategies for instruction and limitations].

Authors:  I Klöpfer-Krämer; P Augat
Journal:  Unfallchirurg       Date:  2010-01       Impact factor: 1.000

Review 2.  Fatigue failure of the cephalomedullary nail: revision options, outcomes and review of the literature.

Authors:  Adam Tucker; Michael Warnock; Sinead McDonald; Laurence Cusick; Andrew P Foster
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-10-17

3.  Case report: unique failure of a Synthes TFNA fenestrated lag screw in a peritrochanteric nonunion.

Authors:  Kevin P Smidt; Kevin J Perry; Lincoln K Andre; Brad J Chauvin; Steve Kautz; Patrick A Massey; R Shane Barton
Journal:  OTA Int       Date:  2022-03-18

4.  Breakage of a Lag Screw of Cephalomedullary Nail: A Technique of Removal.

Authors:  Direk Tantigate; Kongkhet Riansuwan; Banchong Mahaisavariya; Kitichai Sukjaitham
Journal:  Clin Orthop Surg       Date:  2015-05-18

5.  Breakage of a third generation gamma nail: a case report and review of the literature.

Authors:  Takashi Iwakura; Takahiro Niikura; Sang Yang Lee; Yoshitada Sakai; Kotaro Nishida; Ryosuke Kuroda; Masahiro Kurosaka
Journal:  Case Rep Orthop       Date:  2013-05-20

6.  Nonunion with breakage of gamma nail and subsequent fracture in the ipsilateral femur.

Authors:  Takahiro Niikura; Sang Yang Lee; Yoshitada Sakai; Kotaro Nishida; Ryosuke Kuroda; Masahiro Kurosaka
Journal:  Case Rep Med       Date:  2013-02-21

7.  In vitro comparison of resistance to implant failure in unstable trochanteric fractures fixed with intramedullary single screw versus double screw device.

Authors:  Amit Rastogi; Gr Arun; Vakil Singh; Anant Singh; Ashutosh K Singh; Vinay Kumaraswamy
Journal:  Indian J Orthop       Date:  2014-05       Impact factor: 1.251

8.  Intertrochanteric fracture non-unions with implant failure of the gamma nail.

Authors:  Mouhcine Sbiyaa; Adil El Alaoui; Mohammed Admi; Kamal Lahrach; Amine Marzouki; Fawzi Boutayeb
Journal:  Pan Afr Med J       Date:  2016-02-29

9.  Failure of a Rotation Control Gamma 3 Lag Screw Used to Treat a Trochanteric Fracture.

Authors:  Kyungho Choi; Yongtae Kim; Shicheng Zhou; Jihyo Hwang
Journal:  Hip Pelvis       Date:  2018-06-04

10.  Breakage in Two Points of a Short and Undersized "Affixus" Cephalomedullary Nail in a Very Active Elderly Female: A Case Report and Review of the Literature.

Authors:  Giuseppe Rollo; Giuseppe Rinonapoli; Paolo Pichierri; Michele Bisaccia; Auro Caraffa; Luigi Meccariello
Journal:  Case Rep Orthop       Date:  2018-09-13
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.