Abstract
Experiments have already shown that obvious differences exist between the dose distribution of electron beams of a clinical accelerator in a water phantom and the dose distribution of monoenergetic electrons of nominal energy of the clinical accelerator in water, because the electron beams which reach the water surface travelling through the collimation system of the accelerator are no longer monoenergetic. It is evident that, while calculating precisely the dose distribution of any incident electron beams, the energy spectrum of the incident electron beam must be taken into consideration. In this note we shall present a method for determining an effective energy spectrum of clinical electron beams from PDD data (percentage depth dose data). It is well known that there is an integral equation of the first kind which links the energy spectrum of an incident electron beam with PDD through the dose distribution of monoenergetic electrons in the medium, as a kernel function in the integral equation. In this note, the integral equation of the first kind will be solved by using the regularization method. The bipartition model of electron transport will be used to calculate the kernel function, namely the energy deposition due to monoenergetic electron beams in the medium.Entities: