Abstract

To evaluate the results of long Micro Stent II (MS-XL) implantations, 119 MS-XLs were implanted in 102 patients (age, 62.83 years). Nineteen stents (16%) were implanted in saphenous vein grafts; 100 stents (84%) were implanted in native coronary arteries. Twenty-five patients (25%) were treated because of acute myocardial infarction (AMI); 30 patients (29%) because of unstable angina or angina class IV, and 47 patients (46%) because of stable angina. Eighty-six de novo lesions (84%) and 16 restenotic lesions (16%) were treated. Indications for stent implantation include elective, 61 patients (60%); suboptimal balloon angioplasty result, 22 patients (21%); and bailout after balloon angioplasty, 19 patients (19%). Because of residual thrombus after stenting, 27 patients (26%) received abciximab. All patients received ticlopidin for 28 days and acetylsalicylic acid. One hundred and seventeen MS-XLs (98%) were implanted successfully. Additional (shorter) MS-II were implanted in 40 patients (39%). The stented segment length was 45 +/- 20 mm. The minimum lumen diameter increased from 0.5 +/- 0.5 mm before to 2.7 +/- 0.5 mm after stent implantation. The acute gain was 2.2 +/- 0.4 mm. Early clinical events (<4 weeks) include death, 3 (3%); subacute stent thrombosis, 1 (1%); non-Q-wave infarction, 2 (2%); CABG, 1 (1%); vascular complications, 2 (2%). Late clinical events (<6 months) include acute myocardial infarction, 5 (5%); reintervention, 6 (6%); CABG, 1 (1%). The procedural success rate was 88%, and the event free survival at 6 months was 76%. Stenting of long lesions with the MS-XL was successful and associated with an acceptable complication rate. Cathet. Cardiovasc. Intervent. 48:105-112, 1999. Copyright 1999 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  1999        PMID: 10467083     DOI: 10.1002/(sici)1522-726x(199909)48:1<105::aid-ccd22>3.0.co;2-0

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  1 in total

1.  Evaluation and efficacy of long length Pronova XR Bioabsorbable Polymer stent in the treatment of long coronary lesions.

Authors:  M S Hiremath
Journal:  Indian Heart J       Date:  2017-06-23
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.