Body surface potential field representation fidelity: analysis of map estimation procedures.

G Sándor1, G Kozmann, Z Cserjés, N Farkas, I Préda.   

Abstract

The first part of this study analyzed the spatial-temporal error distribution of the Lux-type limited lead system. Quantitative new evidence is reported that the 32-lead anterior subset estimates the further 160 leads with an average amplitude error less than 38.5 microV. The spatial error distribution revealed 8 sites where the error is the highest, primarily on the anterior side, independent of the clinical classification. The second part of the study examined inter-lead-system conversion strategies for interpolating the Lux-192 lead maps from the Montreal-63 data. The methodology based on the Laplacian interpolation yielded an average amplitude error of 143.7 microV and an average correlation of 0.87 for pattern fidelity. In this specific case a modified linear interpolation surpassed the Laplacian method. A presented example illustrates that even in cases when the fidelity of the signal information is heavily compromised the diagnostic information may remain less influenced.

Mesh:

Year:  1999        PMID: 10465568     DOI: 10.1016/s0022-0736(99)90107-3

Source DB:  PubMed          Journal:  J Electrocardiol        ISSN: 0022-0736            Impact factor:   1.438


  1 in total

Review 1.  Body surface potential mapping: historical background, present possibilities, diagnostic challenges.

Authors:  Mihály Medvegy; Gábor Duray; Arnold Pintér; István Préda
Journal:  Ann Noninvasive Electrocardiol       Date:  2002-04       Impact factor: 1.468