Rotary blood pump flow spontaneously increases during exercise under constant pump speed: results of a chronic study.

T Akimoto1, K Yamazaki, P Litwak, K N Litwak, O Tagusari, T Mori, J F Antaki, M V Kameneva, M J Watach, M Umezu, J Tomioka, R L Kormos, H Koyanagi, B P Griffith.   

Abstract

Many types of rotary blood pumps and pump control methods have recently been developed with the goal of clinical use. From experiments, we know that pump flow spontaneously increases during exercise without changing pump control parameters. The purpose of this study was to determine the hemodynamics associated with the long-term observation of calves implanted with centrifugal blood pumps (EVAHEART, Sun Medical Technology Research Corporation, Nagano, Japan). Two healthy female Jersey calves were implanted with devices in the left thoracic cavity. A total of 22 treadmill exercise tests were performed after the 50th postoperative day. During exercise, the following parameters were compared with conditions at rest: heart rate, blood pressure, central venous oxygen saturation (SvO2), pump speed, and pump flow. The pump flow in a cardiac cycle was analyzed by separating the systole and diastole. Compared to the base data, statistically significant differences were found in the following interrelated parameters: the heart rate (66.8 +/- 5.2 vs. 106 +/- 9.7 bpm), mean pump flow (4.8 +/- 0.2 vs. 7.0 +/- 0.3 L/min), and volume of pump flow in diastole (26.0 +/- 1.8 vs. 13.5 +/- 2.5 ml). During exercise, the volume of pump flow in systole was 3 times larger than that measured in diastole. Blood pressure, SvO2, and pump speed did not change significantly from rest to exercise. These results suggested that the mean pump flow depends on the systolic pump flow. Therefore, the increase in the mean pump flow during exercise under constant pump speed was caused by an increase in the heart rate.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10463510     DOI: 10.1046/j.1525-1594.1999.06426.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  13 in total

1.  The future of adult cardiac assist devices: novel systems and mechanical circulatory support strategies.

Authors:  Carlo R Bartoli; Robert D Dowling
Journal:  Cardiol Clin       Date:  2011-11       Impact factor: 2.213

Review 2.  Mechanical circulatory support devices (MCSD) in Japan: current status and future directions.

Authors:  Setsuo Takatani; Hikaru Matsuda; Akihisa Hanatani; Chisato Nojiri; Kenji Yamazaki; Tadashi Motomura; Katsuhiro Ohuchi; Tohru Sakamoto; Takashi Yamane
Journal:  J Artif Organs       Date:  2005       Impact factor: 1.731

3.  Completely pulsatile high flow circulatory support with a constant-speed centrifugal blood pump: mechanisms and early clinical observations.

Authors:  Kenji Yamazaki; Satoshi Saito; Shinichiro Kihara; Osamu Tagusari; Hiromi Kurosawa
Journal:  Gen Thorac Cardiovasc Surg       Date:  2007-04

4.  EVAHEART: an implantable centrifugal blood pump for long-term circulatory support.

Authors:  Kenji Yamazaki; Shinichiro Kihara; Takehide Akimoto; Osamu Tagusari; Akihiko Kawai; Mitsuo Umezu; Jun Tomioka; Robert L Kormos; Bartley P Griffith; Hiromi Kurosawa
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  2002-11

5.  Introduction of fixed-flow mode in the DexAide right ventricular assist device.

Authors:  Diyar Saeed; Alex L Massiello; Shanaz Shalli; Hideyuki Fumoto; Tetsuya Horai; Tomohiro Anzai; Leonard A R Golding; Kiyotaka Fukamachi
Journal:  J Heart Lung Transplant       Date:  2010-01       Impact factor: 10.247

6.  In Vivo Evaluation of a Physiologic Control System for Rotary Blood Pumps Based on the Left Ventricular Pressure-Volume Loop.

Authors:  Joshua Cysyk; Choon-Sik Jhun; Ray Newswanger; Walter Pae; Jenelle Izer; Heidi Flory; John Reibson; William Weiss; Gerson Rosenberg
Journal:  ASAIO J       Date:  2022-12-01       Impact factor: 3.826

7.  Cannula Tip With Integrated Volume Sensor for Rotary Blood Pump Control: Early-Stage Development.

Authors:  Joshua Cysyk; Ray Newswanger; Eric Popjes; Walter Pae; Choon-Sik Jhun; Jenelle Izer; William Weiss; Gerson Rosenberg
Journal:  ASAIO J       Date:  2019 May/Jun       Impact factor: 2.872

Review 8.  The current status of heart transplantation and the development of "artificial heart systems".

Authors:  Martin Strüber; Anna L Meyer; Doris Malehsa; Christiane Kugler; Andre R Simon; Axel Haverich
Journal:  Dtsch Arztebl Int       Date:  2009-07-13       Impact factor: 5.594

9.  Preload-based starling-like control for rotary blood pumps: numerical comparison with pulsatility control and constant speed operation.

Authors:  Mahdi Mansouri; Robert F Salamonsen; Einly Lim; Rini Akmeliawati; Nigel H Lovell
Journal:  PLoS One       Date:  2015-04-07       Impact factor: 3.240

10.  Past and present of cardiocirculatory assist devices: a comprehensive critical review.

Authors:  Gianluca Rigatelli; Francesco Santini; Giuseppe Faggian
Journal:  J Geriatr Cardiol       Date:  2012-12       Impact factor: 3.327

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.